Calculate2:Transform:histogram:range:bwidth Query; Calculate2:Transform:histogram:range:span Command/Query - Iwatsu VOAC760 Manual

Remote control manual
Table of Contents

Advertisement

3.5.3.6.8

CALCulate2:TRANsform:HISTogram:RANGe:BWIDth query

This reads out bin width (width of one bin) in the histogram display.
The bin width of histogram is "whole width / number of bins".
Query syntax
: CALCulate2:TRANsform:HISTogram:RANGe:BWIDth?
Response message <bin_width>
<bin_width> is <NR3 numerical response data> format.
Regardless of the range mode, the bin width of the current histogram display is returned.

3.5.3.6.9 CALCulate2:TRANsform:HISTogram:RANGe:SPAN command/query

This sets and reads out the whole width of the histogram display.
The range of histogram is "center value +/- whole width".
Command syntax
:CALCulate2:TRANsform:HISTogram:RANGe:SPAN {< range_span>|MIN|MAX|DEF}
<range_data>
<Numerical program data>
100E-12 ~ 500E+12
Numerical value beyond limit is set by rounding up maximum value/minimum value, and becomes
execution error.
The value other than specified value is rounded up to large one. No execution error occurs.
The range mode of histogram is automatically set to MANUAL.
Query syntax
:CALCulate2:TRANsform:HISTogram:RANGe:SPAN? [{MIN|MAX}]
Response message <range_data>
<range_data> is <NR3 numerical response data> format.
Regardless of the range mode, the whole width of the current histogram display is returned.
Resolution
Default value
1,2,5 steps
10.0
75

Advertisement

Table of Contents
loading

This manual is also suitable for:

Voac7502

Table of Contents