3.5.3.3 CALCulate1:SCALing subsystem
3.5.3.3.1
CALCulate[1]:SCALing:STATe command/query
This performs setting/reading of ON/OFF for the scaling function.
Command syntax
:CALCulate[1]:SCALing:STATe <off_on>
<off_on>
<Character program data>
OFF
ON
The scaling function and dB calculation function are exclusive functions. When this command and
scaling calculation turn ON, the dB calculation function becomes OFF.
If this command is executed during trigger waiting status or measurement, none is done and the
execution error occurs.
Query syntax
:CALCulate[1]:SCALing:STATe?
Response message <off_on>
<off_on> is <NR1 numerical response data > format.
0 and 1 denote OFF and ON respectively.
3.5.3.3.2
CALCulate[1]:SCALing:MODE command/query
This performs setting/reading of mode setting for the scaling function.
Command syntax
:CALCulate[1]:SCALing:MODE <scale_mode>
<scale_mode>
<Character program data>
SCAL
REC
If this command is executed during trigger waiting status or measurement, none is done and the
execution error occurs.
Query syntax
:CALCulate[1]:SCALing:MODE?
Response message <scale_mode>
<scale_mode> is <Character response data> format.
<Numerical program data>
0
1 (Excluding 0)
Setting
(X-A)*B/C
D/X
62