3.5.8.4.7
[SENSe]:CURRent[:DC]:ZERO:AUTO command/query
This sets and reads out ON/OFF of auto zero function of DCI measurement.
If ONCE is set, auto zero operation is executed once and then OFF is set.
Command syntax
[:SENSe]:CURRent[:DC]:ZERO:AUTO <off_on_once>
<off_on_once>
<Character program data>
OFF
ON
ONCE
If this command is executed during trigger waiting status or measurement, none is done and the
execution error occurs.
Query syntax
[:SENSe]:CURRent[:DC]:ZERO:AUTO?
Response message <off_on_once>
<off_on_once> is <character response data> format.
3.5.8.4.8
[SENSe]:CURRent[:DC]:NULL:STATe command/query
This sets and reads out ON/OFF of null measurement of DCI measurement.
Command syntax
[:SENSe]:CURRent[:DC]:NULL[:STATe] <off_on>
<off_on>
<Character program data>
OFF
ON
Query syntax
[:SENSe]:CURRent[:DC]:NULL[:STATe]?
Response message <off_on>
<off_on> is <NR1 numerical response data> format.
0 and 1 denote OFF and ON respectively.
If this command is executed during trigger waiting status or measurement, none is done and the
execution error occurs.
<Numerical program data>
0
1 (Excluding 0)
<Numerical program data>
0
1 (Excluding 0)
128