3.5.4.2
TRIGger:COUNt command/query
This sets and reads the trigger frequency up to stop of measurement.
Command syntax
:TRIGger:COUNt {<trig_num>|MIN|MAX|DEF}
<trg_num>
<Numerical program data>
1 to 50000 (MAX)
INF (<Numerical program data>)
Numerical value beyond limit is set by rounding up maximum value/minimum value, and becomes
execution error.
The trigger count function is only usable for remote. If the INF is set, it becomes continuous
measurement setting.
In bulk mode, this setting becomes invalid.
If this command is executed during trigger waiting status or measurement, none is done and the
execution error occurs.
Query syntax
:TRIGger:SOURce? [{MIN|MAX}]
Response message <trig_num>
<trig_num> is <NR3 numerical response data> format.
In case INF has been set, NAN(9.91E 37) is returned.
Remark
The frequency of the trigger is a effective setting only in remote state. It is disregarded in the local
state.
Default value
1
93