Configure[:Voltage][:Dc] Command - Iwatsu VOAC760 Manual

Remote control manual
Table of Contents

Advertisement

3.5.6.1

CONFigure[:VOLTage][:DC] command

This sets the measuring function to DCV.
Range and measuring resolution are set with parameter.
Command syntax
:CONFigure[:VOLTage][:DC] [{<range>|MIN|MAX|DEF}[,{<reso>|MIN|MAX|DEF}]]
<range>
<Numerical program data>
100E-3,1,10,100,1000
AUTO(<Character program data>)
This allows unit "V".
Numerical value beyond limit is set by rounding up maximum value/minimum value, and becomes
execution error.
Numerical number other than the specified value is rounded up in larger number. It does not
become execution error.
When the parameter is omitted, the default value is set.
<reso>
<Numerical program data>
range×1ppm to range×10ppm
The setting range/default values differ by range setting.
Numerical value beyond limit is set by rounding up maximum value/minimum value, and becomes
execution error.
Numerical number other than the specified value is rounded up in smaller number. It does not
become execution error.
It is set to the sampling rate responded to the resolution set.
When the parameter is omitted, the default value is set.
For the correspondence to sampling rate, please refer to the correspondence table in the
beginning of SENSe subsystem.
If this command is executed during trigger waiting status or measurement, none is done and the
execution error occurs.
Default value
AUTO
Default value
range1ppm
97

Advertisement

Table of Contents
loading

This manual is also suitable for:

Voac7502

Table of Contents