Calculate2:Transform:histogram:range:mode Command/Query; Calculate2:Transform:histogram:range:center Command/Query - Iwatsu VOAC760 Manual

Remote control manual
Table of Contents

Advertisement

3.5.3.6.6

CALCulate2:TRANsform:HISTogram:RANGe:MODE command/query

This sets and reads out range mode of the histogram display.
Command syntax
:CALCulate2:TRANsform:HISTogram:RANGe:MODE <range_mode>
<range_mode>
<Character program data>
AUTO
MANUAL
FULLSCALE
Query syntax
:CALCulate2:TRANsform:HISTogram:RANGe:MODE?
Response message <range_mode>
<range_mode> is <character response data> format.
3.5.3.6.7

CALCulate2:TRANsform:HISTogram:RANGe:CENTer command/query

This sets and reads out the center value of the histogram display.
The range of histogram is "center value +/- whole width".
Command syntax
:CALCulate2:TRANsform:HISTogram:RANGe:CENTer {<range_center>|MIN|MAX|DEF}
<range_center>
<Numerical program data>
-999.9999E+12 to +999.9999E+12
-999.999E+12 to +999.999E+12
Numerical value beyond limit is set by rounding up maximum value/minimum value within range,
and becomes execution error.
Query syntax
:CALCulate2:TRANsform:HISTogram:RANGe:CENTer? [{MIN|MAX}]
Response message <range_center>
<range_center> is <NR3 numerical response data> format.
Regardless of the range mode, the center value of the current histogram display is returned.
Setting
Auto range
Manual setting
Full scale range
Resolution
Default value
0.000001E-12
0.00001E-12
74
+0.0000E+0
VOAC7602
+0.000E+0
VOAC7502
Models