3.3.8
Instrument specific error occurrence event enable register
This register is used to mask the instrument specific error occurrence event status register.
According to the masked results, the DDE bit of the standard event status register is set.
It is possible to set the mask pattern using the :DDEE command and to read it using the :DDEE[?]
query.
Whether or not the contents of this register are cleared (set to 0) as the power is turned on is set using
the PSC command.
This setting is not affected by the instrument clear nor CLS command.
DDE event register format
7
6
SPE
SFE
AND
AND
0
To DDE bit of the standard event status register
Not used
Not used
5
3
SOE
0
HLE
0
AND
AND
0
0
×
×
L
I
M
OR
RS parity error
RS framing error
RS overrun error
Comparator H < L setting
Scaling calculation over
Not used
1
OVM
0
Instrument specific error event status register
AND
×
O
Instrument specific error event enable
register
L
2
30