Rcl Command (Recall); Sav Command (Save) - Iwatsu VOAC760 Manual

Remote control manual
Table of Contents

Advertisement

3.4.14  RCL command (Recall)
Recalls setup file stored in this instrument's internal memory.
Command syntax
 RCL <register_number_arg>
<register_number_arg> is used to specify a parameter for setting of the setup No. to be recalled
and is a numeric value ranging from "0" to "10" in the <NUMERIC VALUE PROGRAM DATA>
format.
If the numeric value is outside the range, it is discarded and the execution error is given.
A numeric value corresponding to the register No. to be selected for the save/recall operation of
the local operation is specified.
Remarks
The parameter cannot be omitted. If omitted, command error is given.
If an integer value that <register_number_arg> is rounded is outside the range, the execution error
is given and the recall operation is not performed.
If this query is executed during trigger waiting status or measurement, the execution error occurs
and no response is done.
3.4.15  SAV command (Save)
The setup information is saved into the internal memory of this instrument.
Command syntax
 SAV <register_number_arg>
<register_number_arg> is used to specify a parameter for setting of the setup No. to be saved and
is a numeric value ranging from "0" to "10" in the <NUMERIC VALUE PROGRAM DATA> format.
If the numeric value is outside the range, it is discarded and the execution error is given.
A numeric value corresponding to the setup No. to be selected for the save/recall operation of the
local operation is specified.
Remarks
The parameter cannot be omitted. If omitted, command error is given.
If an integer value that <register_number_arg> is rounded is outside the range, the execution error
is given and the save operation is not performed.
If this query is executed during trigger waiting status or measurement, the execution error occurs
and no response is done.
45

Advertisement

Table of Contents
loading

This manual is also suitable for:

Voac7502

Table of Contents