Configure:period Command; Configure:temperature Command - Iwatsu VOAC760 Manual

Remote control manual
Table of Contents

Advertisement

3.5.6.9

CONFigure:PERiod command

This sets the measuring function to FREQ and sets the mode to PERIOD.
Range and measuring resolution are set with parameter.
Command syntax
:CONFigure:PERiod [{<range>|MIN|MAX|DEF}[,{<reso>|MIN|MAX|DEF}]]
<range>
<Numerical program data>
3E-6 to 300E-3
This allows unit "S".
Numerical value beyond limit is set by rounding up maximum value/minimum value, and becomes
execution error.
Numerical number other than the specified value is rounded up in larger number. It does not
become execution error.
When the parameter is omitted, the default value is set.
<reso>
<Numerical program data>
range×0.1ppm to range×100ppm
The setting range/default values differ by range setting.
Numerical value beyond limit is set by rounding up maximum value/minimum value, and becomes
execution error.
Numerical number other than the specified value is rounded up in smaller number. It does not
become execution error.
It is set to the gate time responded to the resolution set.
When the parameter is omitted, the default value is set.
If this command is executed during trigger waiting status or measurement, none is done and the
execution error occurs.
3.5.6.10

CONFigure:TEMPerature command

This sets the measuring function to TEMP.
Temperature sensor type is set with parameter.
Command syntax
:CONFigure:TEMPerature <type>
<type>
<Character program data>
TC
RTD
FRTD
If this command is executed during trigger waiting status or measurement, none is done and the
execution error occurs.
Remark
:CONFigure:TEMPerature command is effective only for VOAC7602, and becomes a command
error for VOAC7502.
Default value
50E-3
Default value
range1ppm
Setting
Thermocouple
Resistance temperature detector 2W
Resistance temperature detector 4W
104