Sense]:Current[:Dc]:Range[:Upper] Command/Query - Iwatsu VOAC760 Manual

Remote control manual
Table of Contents

Advertisement

3.5.8.4.6 [SENSe]:CURRent[:DC]:RANGe[:UPPer] command/query

This sets and reads out the range of DCI measurement.
Command syntax
[:SENSe]:CURRent[:DC]:RANGe[:UPPer] {<range>|MIN|MAX|DEF}
<range>
<Numerical program data>
1E-3
10E-3
100E-3
1
3
This allows unit "A".
Numerical value beyond limit is set by rounding up maximum value/minimum value, and becomes
execution error.
Numerical number other than the specified value is rounded up in larger number. It does not
become execution error.
The auto range setting of DCI measurement is automatically set to OFF.
If this command is executed during trigger waiting status or measurement, none is done and the
execution error occurs.
Query syntax
[:SENSe]:CURRent[:DC]:RANGe[:UPPer]? [{MIN|MAX}]
Response message <range>
<range> is <NR3 numerical response data> format.
A current range set value returns when parameter is not available.
Default value
1
127

Advertisement

Table of Contents
loading

This manual is also suitable for:

Voac7502

Table of Contents