Psc/  Psc? Command/Query (Power On Status Clear) - Iwatsu VOAC760 Manual

Remote control manual
Table of Contents

Advertisement

3.4.10  PSC/  PSC? command/query (Power on Status Clear)
The  PSC command controls the automatic clear of the service request enable register, standard
event status enable register, and instrument-specific event enable register groups at power on.
The values set by the  PSC command can be obtained using the  PSC? query. If the value of the
response data is "0", the status of each enable register is retained even while the power is off. If the
value of the response data is "1", this shows that each of above enable registers is cleared when the
power is turned on again.
Command syntax
 PSC <psc_flag_arg>
<psc_flag_arg> is a parameter for setting of the power on status clear flag and is a numeric value
ranging "  32767" to "  32767" in the <NUMERIC VALUE PROGRAM DATA> format.
When <psc_flag_arg> is set at "0", the status of each enable register is retained even while the
power is turned off. When <psc_flag_arg> is set at "1" (value other than "0"), each enable register
is cleared when the power is turned on again.
Query syntax
 PSC?
Response message <psc_flag>
<psc_flag> is a numeric value showing the contents of the power on status clear flag and is
normally "0" or "1" in the <NR1 NUMERIC RESPONSE DATA> format.
42