Trigger:delay Command/Query; Trigger:slope Command/Query - Iwatsu VOAC760 Manual

Remote control manual
Table of Contents

Advertisement

3.5.4.3

TRIGger:DELay command/query

This sets and reads out the delay time from trigger signal to first sampling.
Command syntax
:TRIGger:DELay {<timer>|MIN|MAX|DEF}
<timer>
<Numerical program data>
0 to 3600
Numerical value beyond limit is set by rounding up maximum value/minimum value, and becomes
execution error.
Numerical values under resolution are rounded up to the nearest whole number.
In bulk mode, this setting becomes invalid.
If this command is executed during trigger waiting status or measurement, none is done and the
execution error occurs.
Query syntax
:TRIGger:DELay? [{MIN|MAX}]
Response message <timer>
<timer> is <NR3 numerical response data> format.
3.5.4.4

TRIGger:SLOPe command/query

This sets and reads out the trigger slope of external trigger (EXT).
Command syntax
:TRIGger:SLOPe <slope>
<slope>
<Character program data>
POS
NEG
If this command is executed during trigger waiting status or measurement, none is done and the
execution error occurs.
Query syntax
:TRIGger:SLOPe?
Response message <slope>
<slope> is <character response data> format.
Resolution
Default value
0.01E-3
0
Setting
Trigger at rising
Trigger at falling
94

Advertisement

Table of Contents
loading

This manual is also suitable for:

Voac7502

Table of Contents