3.5.16.3
STATus:OPERation:ENABle command/query
This sets and reads out the operation event enable register.
Command syntax
:STATus:OPERation:ENABle <status_mask>
<status_mask>
<Numerical program data>
0 to 65535
Numerical value beyond limit is set by rounding up maximum value/minimum value, and becomes
execution error.
In case 1 was set to the bits (bit 0 3 and 6 15) not used for this instrument, it becomes invalid.
However it does not become an execution error.
Query syntax
:STATus:OPERation:ENABle?
Response message <status_mask>
<status_mask> is <NR1 numerical response data> format.
The values from 0 to 65535(16 bits) are repeated. For the description of each bit, please refer to
"3.3 Status report structure."
3.5.16.4
STATus:OPERation:CONDition query
This reads out the operation condition status register.
Query syntax
:STATus:OPERation:CONDition?
The register value by reading is not changed.
Response message <status>
<status> is <NR1 numerical response data> format.
The values from 0 to 65535(16 bits) are repeated. For the description of each bit, please refer to
"3.3 Status report structure."
197