3.4.2 IDN? query (Identification Number)
This reads instrument-specific information for this instrument.
Query syntax
IDN?
Response message<maker_name>,<model_number>,<serial_number>,<software_revision>
The response is configured of four fields separated by commas. The fields are as follows.
<maker_name>
<model_number>
<serial_number>
<software_revision>
Remarks
The response data uses four fields to form the <ARBITRARY ASCII RESPONSE DATA> format.
Therefore, the IDN? query must be the last query in <PROGRAM MESSAGE>. If this rule is
violated, the query error is given and the response message corresponding to the queries after the
IDN? query are not created and sent.
3.4.3 RST command (Reset)
RST command initializes the internal setup to its default values.
The
At this time, however, the settings of the remote interfaces are not changed.
Command syntax
RST
Remarks
If this command is performed during trigger wait status or measurement, the instrument enters the
HOLD mode.
3.4.4 TST? query (Test)
This confirms firmware normality and returns the result.
Query syntax
TST?
If this query is executed during trigger waiting status or measurement, the execution error occurs
and no response is done.
Response message <result>
<result> shows the test results in the <NR1 NUMERIC RESPONSE DATA> format.
Normal case becomes 0.
Abnormal case becomes 1.
Manufacturer
Model No.
Serial No.
Revision No.
38
IWATSU
VOAC7602 (or VOAC7502)
ASCII numeric value (unsigned <NR2>)
ASCII numeric value (unsigned <NR2>)