3.5.16.5

STATus:QUEStionable[:EVENt] query

This reads out the questionable event status register.
Query syntax
:STATus:QUEStionable[:EVENt]?
The register is cleared to 0 after reading.
Response message <status>
<status> is <NR1 numerical response data> format.
The values from 0 to 65535(16 bits) are repeated. For the description of each bit, please refer to
"3.3 Status report structure."
3.5.16.6

STATus:QUEStionable:ENABle command/query

This sets and reads out the questionable event enable register.
Command syntax
:STATus:QUEStionable:ENABle <status_mask>
<status_mask>
<Numerical program data>
0 to 65535
Numerical value beyond limit is set by rounding up maximum value/minimum value, and becomes
execution error.
In case 1 was set to the bits (bit 2  3, 5  8, 10, 13 and 15) not used for this instrument, it becomes
invalid. However it does not become an execution error.
Query syntax
:STATus:QUEStionable:ENABle?
Response message <status_mask>
<status_mask> is <NR1 numerical response data> format.
The values from 0 to 65535(16 bits) are repeated. For the description of each bit, please refer to
"3.3 Status report structure."