3.5.8.5.5
[SENSe]:CURRent:AC:RANGe[:UPPer] command/query
This sets and reads out the range of ACI measurement.
Command syntax
[:SENSe]:CURRent:AC:RANGe[:UPPer] {<range>|MIN|MAX|DEF}
<range>
<Numerical program data>
1
3
This allows unit "A".
Numerical value beyond limit is set by rounding up maximum value/minimum value, and becomes
execution error.
Numerical number other than the specified value is rounded up in larger number. It does not
become execution error.
The auto range setting of ACI measurement is atumatically set to OFF.
If this command is executed during trigger waiting status or measurement, none is done and the
execution error occurs.
Query syntax
[:SENSe]:CURRent:AC:RANGe[:UPPer]? [{MIN|MAX}]
Response message <range>
<range> is <NR3 numerical response data> format.
A current range set value returns when parameter is not available.
Default value
1