Sense]:Resistance:resolution Command/Query; Sense]:Resistance:range:auto Command/Query - Iwatsu VOAC760 Manual

Remote control manual
Table of Contents

Advertisement

3.5.8.6.4

[SENSe]:RESistance:RESolution command/query

The sampling rate of 2 WΩ is set and read out with the measured resolution.
Command syntax
[:SENSe]:RESistance:RESolution {<reso>|MIN|MAX|DEF}
<reso>
<Numerical program data>
range×1ppm to range×10ppm
The range is range set value of 2WΩ.
The setting range/default value differs by the range setting.
For the correspondence to sampling rate, please refer to the correspondence table in the
beginning of SENSe subsystem.
Numerical value beyond limit is set by rounding up maximum value/minimum value, and becomes
execution error.
Numerical number other than the specified value is rounded up in smaller number. It does not
become execution error.
If this command is executed during trigger waiting status or measurement, none is done and the
execution error occurs.
Query syntax
[:SENSe]:RESistance:RESolution? [{MIN|MAX}]
Response message <reso>
<reso> is <NR3 numerical response data> format.
3.5.8.6.5

[SENSe]:RESistance:RANGe:AUTO command/query

This sets and reads out ON/OFF of the auto range of 2WΩ measurement.
Command syntax
[:SENSe]:RESistance:RANGe:AUTO <off_on>
<off_on>
<Character program data>
OFF
ON
If this command is executed during trigger waiting status or measurement, none is done and the
execution error occurs.
Query syntax
[:SENSe]:RESistance:RANGe:AUTO?
Response message <off_on>
<off_on> is <NR1 numerical response data> format.
0 and 1 denote OFF and ON respectively.
Default value
range×1ppm
<Numerical program data>
0
1 (Excluding 0)
140

Advertisement

Table of Contents
loading

This manual is also suitable for:

Voac7502

Table of Contents