3.5.8.3.8
[SENSe]:VOLTage:AC:BANDwidth command/query
This sets and reads out the bandwidth of ACV measurement.
Command syntax
[:SENSe]:VOLTage:AC:BANDwidth {<bandwidth>|MIN|MAX|DEF}
<bandwidth>
<Numerical program data>
20
200
This allows unit "Hz".
Numerical value beyond limit is set by rounding up maximum value/minimum value, and becomes
execution error.
Numerical number other than the specified value is rounded up in smaller number. It does not
become execution error.
If this command is executed during trigger waiting status or measurement, none is done and the
execution error occurs.
Query syntax
[:SENSe]:VOLTage:AC:BANDwidth? [{MIN|MAX}]
Response message <bandwidth>
<bandwidth> is <NR1 numerical response data> format.
3.5.8.3.9
[SENSe]:VOLTage:AC:FREQuency command/query
This sets and reads out ON/OFF of the measured result display of frequency measurement at the
time of ACV measurement.
Command syntax
[:SENSe]:VOLTage:AC:FREQuency <off_on>
<off_on>
<Character program data>
OFF
ON
If this command is executed during trigger waiting status or measurement, none is done and the
execution error occurs.
Query syntax
[:SENSe]:VOLTage:AC:FREQuency?
Response message <off_on>
<off_on> is <NR1 numerical response data> format.
0 and 1 denote OFF and ON respectively.
Default value
20
<Numerical program data>
0
1 (Excluding 0)
122