3.5.4.1

TRIGger:SOURce command/query

This sets and reads the trigger source.
Command syntax
:TRIGger:SOURce <trig_src>
<trg_src>
<Character program data>
IMM
EXT
BUS
In the bulk mode, setting can be made only to IMM. If other than IMM is set, execution error occurs.
In addition, if EXT (external trigger) is set to log stop event in the bulk mode, external trigger input
becomes available automatically.
If this command is executed during trigger waiting status or measurement, none is done and the
execution error occurs.
Query syntax
:TRIGger:SOURce?
Response message <trig_src>
<trig_src> is <character response data> format.
Remark
In a remote mode, an external trigger is assumed to be effective when the trigger source is EXT, and
it is assumed when it is invalid besides. In a remote mode, an external trigger is assumed to be
effective when the trigger source is EXT, and it is assumed when it is invalid besides.
Moreover, the setting of the trigger source is disregarded oppositely in a local mode, and the
movement of an external trigger follows the setting of the TRIG-EXT TRIG menu.
However, the trigger slope setting of an external trigger becomes common in a local mode and a
remote mode.
Setting
Auto trigger
External trigger
Trigger by TRG
92