3.5.3.2

CALCulate1:SMOothing subsystem

The smootiing function can set every each measuring function.
This subsystem performs setting or reading of the smoothing function for measuring function during
startup.
3.5.3.2.1

CALCulate[1]:SMOothing:STATe command/query

This performs setting/reading of ON/OFF for the smooting function.
Command syntax
:CALCulate[1]:SMOothing:STATe <off_on>
<off_on>
<Character program data>
OFF
ON
If this command is executed during trigger waiting status or measurement, none is done and the
execution error occurs.
Query syntax
:CALCulate[1]:SMOothing:STATe?
Response message <off_on>
<off_on> is <NR1 numerical response data > format.
0 and 1 denote OFF and ON respectively.
3.5.3.2.2

CALCulate[1]:SMOothing:POINts command/query

This performs setting/reading of number of sample point for the smoothing function.
Command syntax
:CALCulate[1]:SMOothing:POINts {<point_data>|MIN|MAX|DEF}
<point_data>
<Numerical program data>
2~100
Numerical value beyond limit is set by rounding up maximum value/minimum value within range,
and becomes execution error.
If this command is executed during trigger waiting status or measurement, none is done and the
execution error occurs.
Query syntax
:CALCulate[1]:SMOothing:POINts? [{MIN|MAX}]
Response message <point_data>
<point_data> is <NR1 numerical response data> format figures.
<Numerical program data>
0
1 (Excluding 0)
Resolution
1
61
Default value
10