Sense]:Current[:Dc]:Null:value Command/Query - Iwatsu VOAC760 Manual

Remote control manual
Table of Contents

Advertisement

3.5.8.4.9

[SENSe]:CURRent[:DC]:NULL:VALue command/query

This sets and reads out the null value of DCI measurement.
It is used for the null measurement of DCI measurement.
Command syntax
[:SENSe]:CURRent[:DC]:NULL:VALue {<val_data>|MIN|MAX|DEF}
<val_data>
<Numerical program data>
-999.9999E+12 to +999.9999E+12
-999.999E+12 to +999.999E+12
Numerical value beyond limit is set by rounding up maximum value/minimum value, and becomes
execution error.
This allows unit "A".
Query syntax
[:SENSe]:CURRent[:DC]:NULL:VALue? [{MIN|MAX}]
If this command is executed during trigger waiting status or measurement, none is done and the
execution error occurs.
Response message <val_data>
<val_data> is <NR3 numerical response data> format.
Resolution
Default value
0.000001E-12
+0.000000E+0
0.00001E-12
+0.00000E+0
129
Models
VOAC7602
VOAC7502

Advertisement

Table of Contents
loading

This manual is also suitable for:

Voac7502

Table of Contents