3.5.3.3.3
CALCulate[1]:SCALing:VALA command/query
This performs setting/reading of parameter "A" for the scaling function.
The scaling function is used when mode is "SCAL (X-A) B/C".
Command syntax
:CALCulate[1]:SCALing:VALA {<val_data>|MIN|MAX|DEF}
<val_data>
<Numerical program data>
-999.9999E+12 to +999.9999E+12
-999.999E+12 to +999.999E+12
Numerical value beyond limit is set by rounding up maximum value/minimum value within range,
and becomes execution error.
If this command is executed during trigger waiting status or measurement, none is done and the
execution error occurs.
Query syntax
:CALCulate[1]:SCALing:VALA? [{MIN|MAX}]
Response message <val_data>
<val_data> is <NR3 numerical response data> format figures.
3.5.3.3.4
CALCulate[1]:SCALing:VALB command/query
This performs setting/reading of parameter "B" for the scaling function.
The scaling function is used when mode is "SCAL (X-A) B/C".
Command syntax
:CALCulate[1]:SCALing:VALB {<val_data>|MIN|MAX|DEF}
<val_data>
<Numerical program data>
-999.9999E+12 to +999.9999E+12
-999.999E+12 to +999.999E+12
Numerical value beyond limit is set by rounding up maximum value/minimum value within range,
and becomes execution error.
If this command is executed during trigger waiting status or measurement, none is done and the
execution error occurs.
Query syntax
:CALCulate[1]:SCALing:VALB? [{MIN|MAX}]
Response message <val_data>
<val_data> is <NR3 numerical response data> format figures.
Resolution
Default value
0.000001E-12
+0.0000E+0
0.00001E-12
+0.000E+0
Resolution
Default value
0.000001E-12
+1.0000E+0
0.00001E-12
+1.000E+0
63
Models
VOAC7602
VOAC7502
Models
VOAC7602
VOAC7502