Adjustment By Remote; Remote Adjustment Command/Query List - Iwatsu VOAC760 Manual

Remote control manual
Table of Contents

Advertisement

3.6 Adjustment by Remote

This section explains the command/query to be used for calibrating this instrument. When these
commands are used, select "SCPI" in COMMAND on SYSTEM/REMOTE menu on the main unit.
3.6.1

Remote adjustment command/query List

The following table shows the remote adjustment command/query.
Syntax
:CAL
:ENT
:EXIT <param>
:SAV?
:DCV:ZERO?
:DCV:V100M:POS?
:DCV:V100M:NEG?
:DCV:V1:POS?
:DCV:V1:NEG?
:DCV:V10:POS?
:DCV:V10:NEG?
:DCV:V100:POS?
:DCV:V100:NEG?
:DCV:V1000:POS?
:DCV:V1000:NEG?
:ACV:V100M:ZERO?
:ACV:V100M:FULL?
:ACV:V1:ZERO?
:ACV:V1:FULL?
:ACV:V10:ZERO?
:ACV:V10:FULL?
:ACV:V100:ZERO?
:ACV:V100:FULL?
:ACV:V750:ZERO?
:ACV:V750:FULL?
:ACFC:HZ100K:V100M?
:ACFC:HZ100K:V1?
:ACFC:HZ100K:V10?
:ACFC:HZ100K:V100?
:ACFC:HZ100K:V750?
:DCI:ZERO?
:DCI:A1M:POS?
:DCI:A1M:NEG?
Function
Adjustment subsystem
Shifting to adjustment mode
Shifting from adjustment mode to normal mode
(measurement)
Adjusted value write. Result returning
DCV offset (0V) adjustment
DCV +100mV full scale adjustment
DCV -100mV full scale adjustment
DCV +1V full scale adjustment
DCV -1V full scale adjustment
DCV +10V full scale adjustment
DCV -10V full scale adjustment
DCV +100V full scale adjustment
DCV -100V full scale adjustment
DCV +1000V full scale adjustment
DCV -1000V full scale adjustment
ACV 100mV linearity adjustment 5%
ACV 100mV linearity adjustment Full
ACV 1V linearity adjustment 5%
ACV 1V linearity adjustment Full
ACV 10V linearity adjustment 5%
ACV 10V linearity adjustment Full
ACV 100V linearity adjustment 5%
ACV 100V linearity adjustment Full
ACV 750V linearity adjustment 5%
ACV 750V linearity adjustment Full
ACV 100kHz 100mV frequency characteristic adjustment
ACV 100kHz 1V frequency characteristic adjustment
ACV 100kHz 10V frequency characteristic adjustment
ACV 100kHz 100V frequency characteristic adjustment
ACV 100kHz 750V frequency characteristic adjustment
DCI Offset (0A) adjustment
DCI +1mA full scale adjustment
DCI -1mA full scale adjustment
204
Remark
<param>
"ON" : Adjusted value writing
"OFF": No adjusted value writing
Response 0: Success, -1: Fail

Advertisement

Table of Contents
loading

This manual is also suitable for:

Voac7502

Table of Contents