Memory:state:recall:auto Command/Query; Memory:state:recall:select Command/Query - Iwatsu VOAC760 Manual

Remote control manual
Table of Contents

Advertisement

3.5.13.4

MEMory:STATe:RECall:AUTO command/query

When turning the power supply of this instrument ON, this sets and reads out ON/OFF of the function
which recalls the specified setup.
Command syntax
:MEMory:STATe:RECall:AUTO <off_on>
<off_on>
<Character program data>
OFF
ON
When ON is set, it is started up by setup stored in the internal memory and by the setup specified
at the time of previous completion.
When OFF is set, it is started up by the default setup.
Query syntax
: MEMory:STATe:RECall:AUTO?
Response message <off_on>
<off_on> is <NR1 numerical response data> format.
0 and 1 denote OFF and ON respectively.
If start up is set at the default setup, it deems to be OFF, and in other case, it deems to be ON.
3.5.13.5

MEMory:STATe:RECall:SELect command/query

This sets and reads out the setup to be recalled when the power supply of this instrument is ON.
Command syntax
:MEMory:STATe:RECall:SELect <setup_num>
<setup_num>
<Numerical program data>
0 to 10
Numerical values other than the range become command error.
If 0 is set, it is started up by the setup at the time of previous completion.
Query syntax
:MEMory:STATe:RECall:SELect?
Response message <setup_num>
<setup_num> is <NR1 numerical response data> format.
When the start up by the setup at the time of previous completion has been set, 0 is returned.
<Numerical program data>
0
1 (Excluding 0)
183