3.5.14.6

SETUP:TRANsfer command/query

The command writes the setup data into this instrument, and changes in accordance with the setup
data written into this instrument setting.
The query reads out current setup as the setup data.
This command cannot be used as multi command.
In addition, if this command/query is executed in trigger waiting status or measurement, none is
doone and execution error occurs.
Command syntax
:SETUP:TRANsfer <delimiter><preamble><setup data>
<delimiter>
Refer to each interface section written in "2 Remote control".
<preamble>
#8xxxxxxxx
Xxxxxxxx :ASCII character string of eight digits that shows number of bytes for < setup data >
Insert 0 in the head and make it to eight digits when the number of bytes are less
than eight digits.
It becomes # 800000478 when the number of bytes for < setup data > is 478 bytes.
<setup date>
Setup data of this instrument
The following data can be used.
Data read out by SETUP:TRANsfer:? query
Data stored in USB memory by SETUP:STORe command
Instrument operation: Data stored in USB memory by SETUP-SAVE
Transmit this command according to the procedure of the following 2 steps.
Step 1:
Transmit the SETUP:TRANsfer command without putting the parameter. As a result, it enters
the state that the main unit can receive the setup data (Include preamble)
Step 2:
Transmit < setup data > by putting < preamble > of the above-mentioned format.
Query syntax
:SETUP:TRANsfer?
Response message <preamble><setup data>
This is same format as command.
187