3.5.8.5

SENSe:CURRent:AC subsystem

3.5.8.5.1

[SENSe]:CURRent:AC:NPLCycles command/query

The sampling rate of ACI measurement is set and read out with PLC converted value (sampling
cycle/power supply cycle).
Command syntax
[:SENSe]:CURRent:AC:NPLCycles {<nplc>|MIN|MAX|DEF}
<nplc>
Power supply frequency 50Hz
1
5
20
(DEF)
The setting range differs by the power supply frequency.
For the correspondence to sampling rate, please refer to the correspondence table in the
beginning of SENSe subsystem.
Numerical value beyond limit is set by rounding up maximum value/minimum value, and becomes
execution error.
Numerical number other than the specified value is rounded up in larger number. It does not
become execution error.
When the bandwidth setting of ACI is MID(20Hz~), it is fixed to the default value.
If this command is executed during trigger waiting status or measurement, none is done and the
execution error occurs.
Query syntax
[:SENSe]:CURRent:AC:NPLCycles? [{MIN|MAX}]
Response message <nplc>
<nplc> is <NR3 numerical response data> format.
<Numerical program data>
Power supply frequency 60Hz
1
6
24
(DEF)
130