3.5.8.5.4
[SENSe]:CURRent:AC:RANGe:AUTO command/query
This sets and reads out ON/OFF of the auto range of ACI measurement.
Command syntax
[:SENSe]:CURRent:AC:RANGe:AUTO <off_on>
<off_on>
<Character program data>
OFF
ON
If this command is executed during trigger waiting status or measurement, none is done and the
execution error occurs.
Query syntax
[:SENSe]:CURRent:AC:RANGe:AUTO?
Response message <off_on>
<off_on> is <NR1 numerical response data> format.
0 and 1 denote OFF and ON respectively.
<Numerical program data>
0
1 (Excluding 0)
133