Data:points Query; Data:delete Command; Data:bulk:length Command/Query - Iwatsu VOAC760 Manual

Remote control manual
Table of Contents

Advertisement

3.5.15.4

DATA:POINts query

This reads out number of measured data being stored in LOG data.
Query syntax
:DATA:POINts?
Response message <data_points>
<data_points> is <NR1 numerical response data>format. The values of 0 to 100000 returns.
3.5.15.5

DATA:DELete command

This erases all data in LOG data.
Command syntax
:DATA:DELete
3.5.15.6

DATA:BULK:LENGth command/query

This sets/reads out the memory length for the bulk mode.
Command syntax
:DATA:BULK:LENGth <length>
<length>
<Nemerical program data>
1,000
2,000
5,000
10,000
20,000
50,000
100,000
Any value out of the range is rounded to a maximum / minimum value; resulting in the execution
error.
A numeric value other than the specified value is rounded up. The execution error does not occur.
If this command is executed in trigger waiting status or measurement, none is done; resulting in
execution error.
Query syntax
:DATA:BULK:LENGth? [{MIN|MAX}]
Response message <length>
<length> is the <NR3 numeric value response data> format.
Default value
1000