3.5.8.3.6

[SENSe]:VOLTage:AC:NULL:STATe command/query

This sets and reads out ON/OFF of the null measurement of ACV measurement.
Command syntax
[:SENSe]:VOLTage:AC:NULL[:STATe] <off_on>
<off_on>
<Character program data>
OFF
ON
If this command is executed during trigger waiting status or measurement, none is done and the
execution error occurs.
Query syntax
[:SENSe]:VOLTage:AC:NULL[:STATe]?
Response message <off_on>
<off_on> is <NR1 numerical response data> format.
0 and 1 denote OFF and ON respectively.
3.5.8.3.7

[SENSe]:VOLTage:AC:NULL:VALue command/query

This sets and reads out the null value of ACV measurement.
It is used for the null measurement of ACV measurement.
Command syntax
[:SENSe]:VOLTage:AC:NULL:VALue {<val_data>|MIN|MAX|DEF}
<val_data>
<Numerical program data>
-999.9999E+12 to +999.9999E+12
-999.999E+12 to +999.999E+12
This allows unit "V".
Numerical value beyond limit is set by rounding up maximum value/minimum value, and becomes
execution error.
If this command is executed during trigger waiting status or measurement, none is done and the
execution error occurs.
Query syntax
[:SENSe]:VOLTage:AC:NULL:VALue? [{MIN|MAX}]
Response message <val_data>
<val_data> is <NR3 numerical response data> format.
<Numerical program data>
0
1 (Excluding 0)
Resolution
Default value
0.000001E-12
+0.000000E+0
0.00001E-12
+0.00000E+0
121