3.5.8.9
SENSe:FREQuency subsystem
3.5.8.9.1
[SENSe]:FREQuency:APERture command/query
This sets and reads out the gate time of FREQ measurement.
Command syntax
[:SENSe]:FREQuency:APERture {<time>|MIN|MAX|DEF}
<time>
<Numerical program data>
1E 3
10E 3
100E 3
1.0E0 *
* 1 second of the gating time is effective only for VOAC7602. 100E-3 seconds reaches the
maximum value for VOAC7502.
This allows unit "S".
Numerical value beyond limit is set by rounding up maximum value/minimum value, and becomes
execution error.
Numerical number other than the specified value is rounded up in larger number. It does not
become execution error.
If this command is executed during trigger waiting status or measurement, none is done and the
execution error occurs.
Query syntax
[:SENSe]:FREQuency:APERture? [{MIN|MAX}]
Response message <time>
<time> is <NR3 numerical response data> format.
Default value
100E 3
151