3.5.3.9.6 CALCulate2:TRANsform:METEr:RANGe:LOG:MAX command/ query

It sets/ reads the maximum display value in LOG mode of analog meter display.
Command syntax
:CALCulate2:TRANsform:METEr:RANGe:LOG:MAX {<range_data>|MIN|MAX|DEF}
<range_data>
<Numerical program data>
1.0E-11 ~ 100E+12
The value out of the range is rounded within the range and the execution error occurs.
The setting value to be acquired is only the value of n-th power of 10 (n: integer). If the other value
than this is set, the value is rounded up. In this case, the execution error does not occur.
In addition, the maximum and minimum values in the display range of LOG mode shall satisfy the
relation express below.
1.0E+6 ≧ max. value / min. value ≧ 1.0E+1
If this condition is not satisfied as a result of setting the maximum value, the minimum value is
automatically reset.
The range mode of analog meter display is automatically set to LOG.
Query syntax
:CALCulate2:TRANsform:METEr: RANGe:LOG:MAX? [{MIN|MAX}]
Response message <range_data>
<range_data> is <NR3 numeric value response data> format.
If the range mode is set to the other than LOG, the execution error occurs and 9.91E+37 is
returned.
Resolution
Value of n-th power of 10
(n: integer)
89
Default value
1.0