Data:bulk:post Command/Qury; Data:bulk:event Command/Query - Iwatsu VOAC760 Manual

Remote control manual
Table of Contents

Advertisement

3.5.15.7

DATA:BULK:POST command/qury

This sets/reads out the post-reading count (percent specified) for the bulk mode. It is available when
the stop event is other than NONE.
Command syntax
:DATA:BULK:POST <post>
<post>
<Nemerical program data>
0 ~ 100
Any value out of the range is rounded to a maximum / minimum value; resulting in the execution
error.
If this command is executed in trigger waiting status or measurement, none is done; resulting in
execution error.
Query syntax
:DATA:BULK:POST? [{MIN|MAX}]
Response message <post>
<post> is <NR1 numeric value response data> format.

3.5.15.8 DATA:BULK:EVENt command/query

This sets/reads out the stop event for the bulk mode.
Command syntax
:DATA:BULK:EVENt <event>
<event>
<Character program data>
NONE
EXT
LEVEL
LIMIT
 For "NONE", measurement in the bulk mode does not stop automatically.
Use ABORt command to stop measurement.
If this command is executed in trigger waiting status or measurement, none is done; resulting in
execution error.
Query syntax
:DATA:BULK:EVENt?
Response message < event >
<event > is <Character response data> format.
Resolution
Default value
1
50
Setting
No stop event*
External trigger
Level
Limit calculation