Sense:temperature Subsystem; Sense]:Temperature:nplcycles Command/Query - Iwatsu VOAC760 Manual

Remote control manual
Table of Contents

Advertisement

3.5.8.11

SENSe:TEMPerature subsystem

All command/query below this subsystem is effective only for VOAC7602. It becomes a command error
for VOAC7502.
3.5.8.11.1

[SENSe]:TEMPerature:NPLCycles command/query

The sampling rate of TEMP measurement is set and read out with PLC converted value (sampling
cycle/power supply cycle).
Command syntax
[:SENSe]:TEMPerature:NPLCycles {<nplc>|MIN|MAX|DEF}
<nplc>
Power supply frequency 50Hz
1.67E-3
3.33E-3
6.67E  3
25E  3
50E  3
100E  3
500E  3
1
5
20
(DEF)
The setting range differs by the power supply frequency.
For the correspondence to sampling rate, please refer to the correspondence table in the
beginning of SENSe subsystem.
Numerical value beyond limit is set by rounding up maximum value/minimum value, and becomes
execution error.
Numerical number other than the specified value is rounded up in larger number. It does not
become execution error.
If this command is executed during trigger waiting status or measurement, none is done and the
execution error occurs.
Query syntax
[:SENSe]:TEMPerature:NPLCycles? [{MIN|MAX}]
Response message <nplc>
<nplc> is <NR3 numerical response data> format.
Remark
[:SENSe]:TEMPerature:NPLCycles command/query is effective only for VOAC7602.
It becomes a command error for VOAC7502.
<Character program data>
Power supply frequency 60Hz
2E-3
4E-3
8E  3
30E  3
60E  3
120E  3
600E  3
1
6
24
(DEF)