3.5.9.5
DISPlay:MENU:ANIMation[:STATe] command/query
This sets and reads out ON/OFF of the animation display function of menu display of this instrument
screen.
Command syntax
:DISPlay:MENU:ANIMation <off_on>
<off_on>
<Character program data>
OFF
ON
Query syntax
:DISPlay:MENU:ANIMation?
Response message <off_on>
<off_on> is <NR1 numerical response data> format.
0 and 1 denote OFF and ON respectively.
3.5.9.6 DISPlay:NUMeric:DIGIts command/ query
It sets/ reads the number of display valid digits for the digit mask function in the numeric value display
of PRIMARY display part.
Command syntax
:DISPlay:NUMeric:DIGIts {<digits>|MIN|MAX|DEF}
<digits>
<Numerical program data>
4 to 7
4 to 6
The value out of the range is rounded to the maximum or minimum value and the execution error
occurs.
To set the digits mask function OFF, set the maximum value (7 or 6) to <digits>.
Query syntax
:DISPlay:NUMeric:DIGIts?
Response message <digits>
<digits> is <NR1 numeric value response data> format.
<Numerical program data>
0
1 (Excluding 0)
Default value
7
6
171
Models
VOAC7602
VOAC7502