3.7.7 Measures for sampling rate

The parameters of VOAC752x series to specify the sampling rate are SLOW/MID/FAST.
The relationship between these paramters and this instrument stting is as shown in the table below:
Measurement functions
DCV, DCA, OHM, TEMP
ACV, ACA, O4W
FREQ
DIOD, CONT

3.7.8 About temperature measurement

VOAC7602 supports not only the thermocouple sensor supported by the VOAC752x but also the
resistance temperature detector. To take measures for them, the parameter of the command to set
the temperature sensor is expanded. The range set command of TEMP is used for the temperature
sensor setting.
Parameter list of :TEMP:RANG:VAL[?]
<Numerical program data>
0
1
2
3
4
5
6
7
8
9
Note that when this instrument uses the thermocouple for measurement, it is necessary to set the
cold contact temperature using :TEMP:JUNC command.
And, VOAC7502 does not support temperature measurement.
Parameters
SLOW|MID|FAST
For LINE FREQ: 50Hz
SLOW: 20PLC, MID: 0.5PLC, FAST: 6.67mPLC
For LINE FREQ: 60Hz
SLOW: 24PLC, MID: 0.6PLC, FAST: 8mPLC
SLOW|MID|FAST
For LINE FREQ: 50Hz
SLOW: 20PLC, MID: 5PLC, FAST: 1PLC
For LINE FREQ: 60Hz
SLOW: 24PLC, MID: 6PLC, FAST: 1PLC
SLOW|MID|FAST
GATE TIME value is set or acquired.
VOAC7602:SLOW: 1s、MID: 100ms、FAST: 10ms
VOAC7502:SLOW: 100ms、MID: 10ms、FAST: 1ms
Always MID
For LINE FREQ: 50Hz
MID: 0.5PLC
For LINE FREQ: 60Hz
MID: 0.6PLC
<Character program data>
TCK
TCR
TCK
TCT
TCJ
TCE
R2P
R2J
R4P
R4J
If BAND setting of ACV or ACA is 20,
MID or FAST results in execution error.
Execution error for command
Temperature sensor
K thermocouple
R thermocouple
K thermocouple
T thermocouple
J thermocouple
E thermocouple
2-wire Pt 100 resistance temperature detector
2-wire JPt 100 resistance temperature detector
4-wire Pt 100 resistance temperature detector
4-wire JPt 100 resistance temperature detector