Patient Auxiliary Leakage: Connection to DUT
Figure 2-15b illustrates the connections between the Guardian 6100 Plus, G30 corded
product adapter, G32 isolation transformer and device under test for a Patient Auxiliary
Leakage current test.
G30
Corded
Product
Adaptor
*
DRIVE+ and DRIVE- Output Terminals
Add bolt and screw SPADE
between bolt and second
connector
SPADE lug
For High Current Ground Bond Testing, make the connection to the DRIVE+ and
DRIVE- terminals with the spade lug behind the nut. Use the Bushing Driver Tool
Page 92 of 177
www.valuetronics.com
QuadTech
Plug into G30
20A
125V
DO NOT TOUCH WHILE
UNDER TEST
Patient Leads
MAX.
VOLTAGE
5kVAC
6kVDC
GB FLOATING
GPIB
SCAN
RS232
SENSE
DRIVE
HV OUTPUT
HIGH VOLTAGE
MAX 5kVAC
6kVDC
N-OUT (CH-1)
L-OUT (CH-1)
Figure 2-15b: Patient Auxiliary Leakage Test using the G6100 Plus
included to secure the nut as shown in Figure 2-15b.
Device UnderTest
DUT
G15
Red
Lead
PE
UNDER
WARNING
!
TEST
CAUTION
!
PASS
UNDER
TEST
PASS
100V
120V
FAIL
220V
REMOTE
REMOTE
240V
OUTPUT
INPUT
PATIENT PROBE
6000-07
#1
#2
HIPOT / LINE LEAKAGE
SCANNER
FUSE RATING
T20A / 250V
NOTE
150799 Rev A3
G32: 1000VA Isolation Transformer
ISOLATING OUTPUT
LINE INPUT VOLTAGE
DUT POWER
SELECTOR
90V - 132V~
T10.0A 250V
198V - 250V~
T5.0A 250V
115V~/230V~
50/60Hz 1200VA MAX
SINGLE PHASE
RESET
START
FAN
o
50
C AUTO
ON/OFF
FAIL
START
RESET
COM
INTER
LOCK
LINE INPUT
INPUT FUSE
N-IN
L-IN
Operation
FUSE RATING
100/120V T8A 250V
220/240V T4A 250V
Need help?
Do you have a question about the Guardian 6100 Plus and is the answer not in the manual?