To Enable Calibration:
The Guardian 6100 Plus instrument requires a minimum 30-minute warm up period prior
to calibration. With the instrument in standby status ([STOP] button previously pressed
and no lights flashing) remove the front panel calibration seal and push (using tip of
small screw driver) the recessed switch through the hole in the front panel labeled Cal-
Enable (to the in position).
From Menu display:
Select item 3 "CALIBRATION" by Pressing [3] [ENTER] Display reads "password".
Press [7] [9] [3] [1] Enter
Select [DEVICE]
4.3.1.1
Connect the OUTPUT terminal of the Guardian 6100 Plus to the input terminal of the
AC/DC high voltage meter. Connect the RTN/LOW terminal of the Guardian to the
GND terminal of the voltmeter. Set the voltmeter to AC and 200V range.
Press [UP] to go to CAL step 1. (ACV 5kV Offset
Press [STOP]
Press [START]
Using keypad enter the reading from the voltmeter followed by the [ENTER].
Press [STOP] to accept reading.
Press [UP] to go to CAL step 2. (ACV 5kV Full
Change the range on the voltmeter to 20kV.
Press [START]
Using keypad enter the reading from the voltmeter followed by the [ENTER].
Press [STOP] to accept reading.
4.3.1.2
Change the setting on the Valhalla to DC and 200V range.
Press [UP] to go to CAL step 3. (DCV 6kV Offset
Press [STOP]
Press [START]
Using keypad enter the reading from the voltmeter followed by the [ENTER].
Press [STOP] to accept reading.
Press [UP] to go to CAL step 4. (DCV 6kV Full
Change the range on the voltmeter to 20kV.
Press [STOP]
Press [START]
Using keypad enter the reading from the voltmeter followed by the [ENTER].
Press [STOP] to accept reading.
4.3.1.3
Service & Calibration
www.valuetronics.com
AC Voltage Calibration
DC Voltage Calibration
IR Voltage Calibration
100V)
4.000kV)
100V)
4.000kV)
150799 Rev A3
Page 167 of 177