AC Hipot Programming – continued from last page.
Press DOWN arrow [⇓]
STEP SETTING
01.
Test Step
02.
Test Mode
03.
Voltage
04.
High Limit
05.
Low Limit
06.
Arc Limit
07.
Arc Filter
08.
Test Time
09.
Ramp Time
10.
CHNL (H-L)
0 - 999 sec
Press [3][.][0][ENTER] to set ramp time = 3.0s
Press DOWN arrow [⇓]
Display reads: "10. CHNL (H-L) : OFF"
Scanner #1 is the internal 6000-07 Hipot/Leakage Current Scanner. In AC mode, only
Channels 1 and 3 are available on Scanner #1. CH1 can be set to H or X. CH3 can be set
to H, L or X.
Press the function key [SETUP] to enter scanner setup menu.
STEP SETTING
01.
Test Step
02.
Test Mode
03.
Voltage
04.
High Limit
05.
Low Limit
1
06.
Arc Limit
07.
Arc Filter
X
08.
Test Time
09.
Ramp Time
10.
CHNL (H-L)
PRESS NUMBER KEYS
Press [1] to set Channel 1 to H (High); Press [3] twice to set Channel 3 to L (Low)
Press the function key [EXIT] to return to AC program menu.
Press DOWN arrow [⇓]
The backlit box is now again at the top of the list adjacent to "Test Step: ___3___ ". STOP.
To continue programming the 7-step example:
Press the function key [UP] to select Test Step = 4. Instructions continued in ¶
OR
To End Programming and perform an AC test:
Press [MENU] to exit program mode.
Perform offset (¶2.7.1), perform Get Cs(¶2.7.2), connect DUT (¶2.7.3) then press green
[START] button.
Page 64 of 177
www.valuetronics.com
3
:
:
AC
:
1.500kV
0.500mA
:
OFF
:
OFF
:
3-230 kHz
:
3.0s
:
OFF
:
OFF
:
0 = OFF
Remote
Lock
3
:
:
AC
1.500kV
:
:
0.500mA
SETUP SCANNER - 1
OFF
:
2
3
4
5
OFF
6
7
8
:
3-230 kHz
:
X
X
X
X
X
X
X
3.0s
:
OFF
:
OFF
:
Remote
Lock
Offset
Error
NEXT BOX
EXIT
Offset
Error
150799 Rev A3
Use Numeric Keys then [ENTER]
Enter the Ramp Time in seconds (s)
0 - 999 seconds;
0 = OFF
Press Numeric Keys then [ENTER]
Use the Numeric Key that corresponds to the Channel #
The Numeric Key acts as a toggle switch: X, H or L
Program the scanner channels
X = Not used
H = High (High voltage/current applied to this channel)
L = Low (This channel is the return)
Press Function Key [NEXT BOX] to setup next scanner
Press Function Key [EXIT] to exit scanner setup