QuadTech Guardian 6100 Plus Instruction Manual page 51

Electrical safety analyzer

Advertisement

Step
1
Mode
GB
3
I = 30A
4
HL = 100mΩ
5
LL = Off
6
Test: 3s
7
Twin Port: Off
8
CHNL: OFF
9
10
11
If the multi-step example of Table 2-6 is followed, then the resultant test display will be
as illustrated Figure 2-6. Step-by-step program instructions are in paragraphs 2.4.1
through 2.4.7 and test instructions are in paragraph 2.7.
Operation
www.valuetronics.com
Table 2-6: Multi-Step Program Example
2
3
OSC
AC
Open: 50%
V= 1.5kV
Short: 300%
HL= 0.5mA
CHNL: X
LL = Off
Arc: Off
Fil: 3-230kHz
Test: 3s
Ramp: 3s
CHNL: X
TEST - M003 - EXAMPLE
MODE
SOURCE
100.0m Ω
01
GB
30.0A
02
OSC
Open
50%
03
AC
1.500kV
0.500mA
04
PA
CHANGE LEAD
05
DC
2.400kV
0.500mA
10.0G Ω
06
IR
1.000kV
07
LC
0.300kV
10.00mA
Figure 2-6: Multi-Step Example Test Screen
150799 Rev A3
4
5
PA
DC
Change Lead
V= 2.4kV
UT: On
HL= 0.5mA
LL = Off
Dwell: 3s
Arc: Off
Fil: 3-230kHz
Test: 3s
Ramp: 3s
CHNL: X
OFFSET
LIMIT
RES.
GET Cs
PAGE UP
PAGE DOWN
SCANNER - 1
1 2 3 4 5 6 7 8
GB
Remote
Lock
Offset
Error
6
7
IR
LC
V = 1kV
UL2601
LL=10000MΩ
Normal
HL = Off
GB Sw: Open
Test: 3s
M: L-G
Ramp: 3s
HL = 6mA
CHNL: X
LL = Off
P=V
PH = 300V
PL = 0V
Test: 3s
CHNL:
Page 51 of 177