Download Print this page

QuadTech Guardian 6100 Plus Instruction Manual page 140

Electrical safety analyzer

Advertisement

:RESult
:ALL
HI
LO
ARC
IO
ADV
OVER
ADI OVER
OSC SHORT (DEC) = 103; OSC OPEN (DEC) = 98
:RESult
Page 140 of 177
www.valuetronics.com
:JUDGment?
Command: :SOURce : SAFEty : RESult : ALL : JUDGment?
This command queries all the judgment results. The return format is:
PASS
USER STOP
CAN NOT TEST
TESTING
STOP
HEX and DEC values:
GB MODE
HEX
DEC
11
17
12
18
---
---
---
---
16
22
17
23
:OMETerage?
Command: :SOURce : SAFEty : RESult : ALL : OMETerage?
This command queries all the voltage (output meter) readings for the step.
:MMETerage?
Command: :SOURce : SAFEty : RESult : ALL : MMETerage?
This command queries all the Measure Meter readings for the step.
:TIME :TEST?
Command: :SOURce : SAFEty : RESult : ALL : TIME : TEST?
This command reads back all step test times including failure times.
:LAST :JUDGment?
Command: :SOURce : SAFEty : RESult : LAST : JUDGment?
This command queries status while testing.
:COMPleted?
Command: :SOURce : SAFEty : RESult : COMPlete?
This command queries if all the test results are completed.
The return format is a "1" or a "0".
:AREPort <Boolean> ON⏐OFF (RS232 Interface only)
Command: :SOURce : SAFEty : RESult : AREPort
This command sets the Automatic Reporting of the test results.
74
71
72
73
70
AC MODE
DC MODE
HEX
DEC
HEX
21
33
31
22
34
32
23
35
33
24
36
34
26
38
36
27
39
37
150799 Rev A3
116
113
114
115
112
IR MODE
DEC
HEX
49
41
50
42
51
---
52
---
54
46
55
47
DEC
65
66
---
---
70
71
Interface

Advertisement

loading
Need help?

Need help?

Do you have a question about the Guardian 6100 Plus and is the answer not in the manual?