Summary of Contents for QuadTech Guardian 6100 Plus
Page 1
The material in this manual is for informational purposes only and is subject to change, without notice. QuadTech assumes no responsibility for any error or for consequential damages that may result from the misinterpretation of any procedures in this publication.
Page 2
Page 2 of 177 150799 Rev A3 www.valuetronics.com...
Page 4
Contents (Continued) Operation - Section 2 - Continued MENU – Continued 2.3.6 FAIL LOCK ..................47 2.3.7 ERROR LOG ..................47 PROGRAM ....................49 2.4.1 Programming a Ground Bond (GB) Test...........53 2.4.1.1 Twin Port ................56 2.4.2 Open/Short Circuit (OSC) Mode ............57 2.4.3 Programming an AC Hipot Test ............61 2.4.4 Pause (PA) Mode ................65 2.4.5 Programming a DC Hipot Test ............67 2.4.6 Programming an Insulation Resistance (IR) Test ......72...
Page 6
Page 6 of 177 150799 Rev A3 www.valuetronics.com...
Page 7
QuadTech's applicable published specifications. If within one (1) year after original shipment it is found not to meet this standard, it will be repaired, or at the option of QuadTech, replaced at no charge when returned to a QuadTech service facility.
Page 8
Page 8 of 177 150799 Rev A3 www.valuetronics.com...
Page 9
Specifications Dielectric Strength Range: 0.5 to 5kV AC, in 1V steps AC Output Voltage: ± (1% of setting +5V) Regulation: Frequency: 50-600Hz, in 10Hz increments ± (1% of reading +5V) Accuracy: Voltage Display: Resolution: 1V steps AC Current Display: Range: 1µA to 40mA AC, in 1µA or 10µA steps ±...
Page 11
Specifications (Continued) ARC Detection Range: 1mA – 20mA AC, 1mA – 12mA DC Detection Current: Pulse Width: Minimum: 40µs, 20µs, 10µs or 4µs Open/Short Circuit Detection Mode Detection Mode: <100V, 600Hz Check for Open & Short against a standard capacitance value, Cs General Features Limits: Hi/Lo Programmable during Test Time...
Page 12
Instruction Manual AC Power Cable Supplied: • • Calibration Certificate HV Lead Set • • Power Entry Adapter Cable GC Lead Set Description Catalog No. Ordering Safety Analyzer Guardian 6100 Plus Information: Page 12 of 177 150799 Rev A3 www.valuetronics.com...
Page 13
Accessories Accessories Included Item Quantity QuadTech P/N AC Power Cord 4200-0300 Power Line Fuse 8A 250V SB 520157 Power Line Fuse 4A 250V SB 520053 Fuse for 6000-07 Scanner, 20A 250V SB 520124 High Voltage Lead Set, 1m with alligator clips...
Page 14
Safety Precautions WARNING The Guardian 6100 Plus Electrical Safety Analyzer can provide an output voltage as high as 6000VDC (5000VAC) to the external device under test (DUT). Although the Guardian unit is designed with full attention to operator safety, serious hazards could occur if the instrument is used improperly and these safety instructions are not followed.
Page 15
Safety Symbols The product is marked with the following safety symbols. Product will be marked with this symbol (ISO#3864) when it is necessary for the user to refer to the instruction manual in order to prevent injury or equipment damage. Product marked with this symbol (IEC417) indicates presence of direct current.
Page 16
Refer to paragraph 1.4.3 to change a fuse and to change the line voltage selector. Connect the Guardian 6100 Plus unit AC power cord to the source of proper voltage. The Guardian 6100 Plus is designed to be operated with its chassis connected to earth ground.
Page 17
Test parameters must be set before the G6100 Plus unit can be zeroed. Zeroing/Offset After setting your test parameters, zero the Guardian 6100 Plus unit by using the automatic offset. With no device connected, connect the appropriate cable (or other fixture) into the front panel OUTPUT connectors.
Page 18
Figure COI-1 illustrates the OPEN configuration of the S02 Leads for an AC Hipot, DC Hipot or Insulation Resistance test. QuadTech Guardian 6100 Plus STOP Electrical Safety Analyzer Twin...
Page 19
AC Hipot Measurement Example: Turn Power ON. Let Guardian 6100 Plus unit warm-up 15 minutes. Connect Black ground cable to G6100 Plus DRIVE- terminal. Connect Red high voltage cable to G6100 Plus OUTPUT terminal.
Page 21
Product Overview The Guardian 6100 Plus Safety Analyzer combines six critical safety tests into a single instrument, these being AC hipot, DC hipot, insulation resistance, leakage current, ground bond and Open/Short detection. The hipot test can be programmed over a voltage range of 0.05 to 5kV AC and 0.05 to 6kV DC with a min/max leakage current...
Page 22
Figure 1-1 illustrates the controls and indicators on the front panel of the Guardian 6100 Plus. Figure 1-2 is a detailed illustration of the key pad (5 of Figure 1-1). Table 1-1 identifies the controls and indicators with descriptions and functions. QuadTech Guardian 6100 Plus STOP Twin Electrical Safety Analyzer...
Page 23
Table 1-1 : G6100 Plus Front Panel Controls and Indicators Reference Name Type Function Number Figures 1-1 and 1-2 POWER White Toggle Switch Applies AC power to unit, 0=OFF, 1=ON Display 320x240 LCD Indicates step, mode, limits, test setup, test result Cal Enable Recessed switch Used by Qualified Service Personnel for unit calibration.
Page 24
Figure 1-3 illustrates the controls and indicators on the rear panel of the Guardian 6100 Plus. Figure 1-4 is a detailed illustration of the two remote connsctors on the rear panel of the Guardian 6100 Plus. Table 1-2 identifies them with description and function. GB FLOATING...
Page 25
Table 1-2: G6100 Plus Rear Panel Connectors and Controls Reference Name Type Function Number Figure 1-3 6000-07 Internal Scanner Hipot/Line Leakage Current Scanner N-OUT CH-1 White Banana Plug Neutral Output to DUT L-OUT CH-1 White Banana Plug Line Output to DUT Red Banana Plug Patient Probe #1 Black Banana Plug...
Page 26
An open space of at least 75mm is recommended behind the rear panel. 1.4.3 Power Requirements The Guardian 6100 Plus can be operated from a power source of 90 to 130 VAC or 200 to 250 VAC. Power connection is via the rear panel through a standard receptacle.
Page 27
1.4.4 Safety Inspection Before operating the instrument inspect the power inlet module on the rear of the Guardian 6100 Plus to ensure that the properly rated fuse is in place, otherwise damage to the unit is possible. Refer to paragraph 1.4.3.
Page 29
Section 2 : Operation Terms and Conventions Table 2-1 : Measurement Unit Prefixes Multiple Scientific Engineering Symbol 10 15 1000000000000000 Peta 10 12 1000000000000 Tera 10 9 1000000000 Giga 10 6 1000000 Mega 10 3 1000 Kilo 10 -3 .001 milli 10 -6 .000001...
Page 30
Discharge: The act of draining off an electrical charge to ground. Devices that retain charge should be discharged after an IR or DC HiPot test. Frequency: The rate at which current or voltage reverses polarity and then back again completing a full cycle, measured in Hertz (Hz) or cycles/second.
Page 31
Leakage Current (LC): Leakage Current: The residual flow of current that flows through the insulation after a high voltage has been applied for a period of time. The leakage current is equal to the applied voltage divided by the insulation resistance. Leakage current is the main measured value for AC hipot and DC hipot.
Page 32
The gradual increase or decrease of voltage or current over a period of time (step). Step: The Guardian 6100 Plus can perform up to 50 tests in a sequence. The step number indicates in which order the tests will be performed. For example if step 1 is a ground...
Page 33
15 minutes prior to use. WARNING DO NOT TURN INSTRUMENT POWER ON OR OFF WITH TEST DEVICES CONNECTED. Note: Read the entire manual before operating the Guardian 6100 Plus instrument. Operation 150799 Rev A3 Page 33 of 177 www.valuetronics.com...
Page 34
MAIN MENU When powered ON, the G6100 instrument will boot to the MAIN MENU display. At any time, press [MENU] to return to the MAIN MENU display. Within the MAIN MENU there are 7 vertical sub-menus: System, Option, Calibration, Key Lock, New Security Code, Fail Lock and Error Log.
Page 35
2.3.1 SYSTEM To access SYSTEM SETUP, press the [MENU] function key at the top left hand corner of the keypad and press the numerical key [1]. Within SYSTEM SETUP there are 11 parameters: Contrast, Beeper Volume, Compensate, DC 50V AGC, Discharge Voltage, Pass ON, Use Source, After Fail, AC Offset, LC Offset and IEEE-488.1.
Page 36
2.3.1.1 Contrast To access Contrast, press [MENU], [1] and the [↓] or [↑] arrow on the keypad so the backlit box is to the right of Contrast. Use the function keys [UP] or [DOWN] to increase or decrease the contrast. The display Contrast is adjustable from 1 – 31. value of 1 is brightest , a value of 31 is darkest.
Page 37
2.3.1.3 Compensate This setting is valid for Guardian 6100 Plus line leakage test only. Compensate is the LC input voltage compensation in which a correction (% value) can be added to better simulate what the true line leakage would be. The simulation correction value can be programmed in whole number increments of ±(5-50%).
Page 38
2.3.1.5 Discharge-V The Discharge-V function lets the user set the voltage that the G6100 Plus instrument will discharge to upon completion of a hipot test. The time it will take to discharge to a specific voltage is dependent on the capacitance of the DUT. The discharge time is a natural log function.
Page 39
2.3.1.7 Use Source This function is valid for the Guardian 6100 Plus line leakage test only. The Use Source function is used when the G6100 Plus instrument is connected to a Chroma 6400 Series AC Power Source. It is valid in simulated leakage current mode only and requires an RS- 232 connection to the 6400.
Page 40
When “Continue on Fail” is selected, the Guardian 6100 Plus will continue to run any remaining steps that are programmed after the step in which the Fail occurred. The Guardian 6100 Plus will then indicate a failure at the end of the test.
Page 41
2.3.1.10 LC Offset This function is valid for the Guardian 6100 Plus line leakage test only. If the measured offset value is higher than the programmed LC Offset value then the Displayed Current = (Measured value) – (Offset value). If the measured offset is lower than the LC offset value then the Displayed Current = the square root of (Measured value) –...
Page 42
2.3.1.11 IEEE-488.1 Select IEEE-488.1 to ON to make the remote commands compatible with the QuadTech Guardian 6000 instrument. When IEEE-488.1 is set to ON, only the Guardian 6000 commands are valid. Refer to the Guardian 6000 instruction manual (P/N 150354) for a set of commands.
Page 43
2.3.2.1 RS-232 Use RS-232 to set the baud rate, parity and flow control for the RS-232 interface. To access RS-232 Setup, enter the Option Menu by pressing [MENU] and [2] then the function key [RS-232]. The backlit box is next to ‘Select baud rate’. Use the [UP] or [DOWN] function key to select the baud rate = 300, 600, 1200, 2400, 4800, 9600 or 19200.
Page 44
If the optional PRINTER interface is installed, use PRINT to select when and what the Guardian 6100 Plus instrument should print. To access Print Setup, enter the Option menu by pressing [MENU], [2] and then the function key [PRINT]. The backlit box is next to ‘Print On’.
Page 45
2.3.3 CALIBRATION Refer to Section 4, Sevice & Calibration for the calibration procedure. Calibration of the Guardian 6100 Plus is recommended on an annual basis. The instrument can be calibrated by a qualified service person if traceable calibration equipment and standards are available.
Page 46
TEST function is operational. Use Fail Lock when you want a supervisor to be notified when a failure occurs. Once a failure occurs, no test can be run until the password is entered. When powered is OFF, the Guardian 6100 Plus will boot to the TEST menu when FAIL LOCK is ON.
Page 47
MAIN MENU MEMORY SYSTEM OPTION PRESET CALIBRATION KEY LOCK PROGRAM NEW SECURITY CODE FAIL LOCK ERROR LOG TEST Remote Lock Offset Error PRESS to view contents of Error Log MAIN MENU SYSTEM Error Message - 1 -102 Syntax error ERROR LOG EXIT Remote Lock...
Page 48
-400 Queue error Output buffer is out of range -410 Query Interrupted No read command following a query command -420 Query Unterminated No data in queue and device requested to read queue PROGRAM To access the PROGRAM menu, press [MENU] then press the function key to the right of the display that corresponds to PROGRAM.
Page 49
MAIN MENU MEMORY SYSTEM OPTION PRESET CALIBRATION KEY LOCK PROGRAM Enter Program Mode NEW SECURITY CODE FAIL LOCK ERROR LOG TEST Remote Lock Offset Error STEP SETING Test Step Test Mode Current High Limit 100.0mΩ DOWN Low Limit Test Time 3.0s Twin Port CHNL (H-L)
Page 50
Table 2-5: Program Parameters AC Hipot DC Hipot Test Step Test Step Test Step Test Step Test Step Test Step Test Step 1-50 1-50 1-50 1-50 1-50 1-50 1-50 Test Mode Test Mode Test Mode Test Mode Test Mode Test Mode Test Mode Current Voltage...
Page 51
Table 2-6: Multi-Step Program Example Step Mode I = 30A Open: 50% V= 1.5kV Change Lead V= 2.4kV V = 1kV UL2601 HL = 100mΩ Short: 300% HL= 0.5mA UT: On HL= 0.5mA LL=10000MΩ Normal LL = Off CHNL: X LL = Off LL = Off HL = Off...
Page 52
Floating Ground: The HV Output terminal on the rear panel of the Guardian 6000 Plus is used to make a floating ground connection to the device under test. It is referred to as Channel 3 and is programmed H (High), L (Low) or X (Off) in the last step of OSC, AC, DC and IR modes.
Page 53
2.4.1 Programming a Ground Bond (GB) Test In a GB test, high AC current is applied between a conductive surface of the DUT and ground connection. The ground circuit resistance is then calculated and displayed. A high limit is required in a GB test. The high limit is the upper value for the test to be considered a PASS.
Page 54
GB Programming – continued from last page. Press DOWN arrow [⇓] Use Numeric Keys then [ENTER] STEP SETTING Test Step Enter the Test Current in amperes (A) Test Mode Current 1 - 60A, 0 = OFF High Limit 100.0mΩ Low Limit 1 - 30A Test Time 3.0s...
Page 55
GB Programming – continued from last page. Press DOWN arrow [⇓] Use Numeric Keys then [ENTER] STEP SETTING Test Step Enter the Test Time in seconds (s) Test Mode Current 30.0A 0, 0.3 - 999s 0 = Continuous High Limit 100.0mΩ...
Page 56
GB Programming – continued from last page. NOTE: Scanner #1 is the 6000-07 Leakage Current Scanner. It can not be programmed for GB. Press the function key [SETUP] to enter scanner setup menu (For External Scanner). Use Numeric Keys then [ENTER] STEP SETTING Use the Numeric Key that corresponds to the Channel # Test Step...
Page 57
2.4.2 Open/Short Circuit (OSC) Detection Mode The Open/Short Circuit Detection Mode will ensure the Device Under Test is connected properly and does not have a short circuit. In Program mode OSC can be selected as one of the steps within the test. There are two programmable parameters Open % and Short %.
Page 58
Back to programming OSC mode. If not already in PROGRAM mode: Press [MENU] to access the PROGRAM menu Press the function key [PROGRAM] If continuing example: Press the function key [UP] to select Step # = 2 Use Function Key STEP SETTING Toggle UP: 1 - 50 Test Step...
Page 59
OSC Programming – continued from last page. Press DOWN arrow [⇓] Use Numeric Keys then [ENTER] STEP SETTING Test Step Enter the Open comparison value in percent (%) Test Mode Open Chk. 10% - 100%, default = 50% Short Chk. 300% CHNL (H-L) 10% - 100%...
Page 60
OSC Programming – continued from last page. Scanner #1 is the internal 6000-07 Hipot/Leakage Current Scanner. In OSC mode, only Channels 1 and 3 are available on Scanner #1. CH1 can be set to H or X. CH3 can be set to H, L or X.
Page 61
If continuing example: Press the function key [UP] or [DOWN] to select Step # = 3 Use Function Key STEP SETTING Toggle UP: 1 - 50 Test Step Test Mode Current 100.0m Ω High Limit DOWN Toggle DOWN: 50 - 1 Low Limit Test Time 3.0s...
Page 62
AC Hipot Programming – continued from last page. Press DOWN arrow [⇓] Use Numeric Keys then [ENTER] STEP SETTING Enter the Test Voltage in kilo-volts (kV) Test Step Test Mode Voltage 0.05 - 5 kV (50 - 5000 V) High Limit 0.500mA Low Limit Arc Limit...
Page 63
AC Hipot Programming – continued from last page. Press DOWN arrow [⇓] Use Numeric Keys then [ENTER] STEP SETTING Enter the Arc Limit in milli-amps (mA) Test Step Test Mode 1.500kV Voltage 0 - 20 mA, 0 = OFF High Limit 0.500mA Low Limit Arc Limit...
Page 64
AC Hipot Programming – continued from last page. Press DOWN arrow [⇓] Use Numeric Keys then [ENTER] STEP SETTING Enter the Ramp Time in seconds (s) Test Step Test Mode Voltage 1.500kV 0 - 999 seconds; 0 = OFF 0.500mA High Limit Low Limit Arc Limit...
Page 65
2.4.4 Pause (PA) Mode “PAUSE” is a mode selection that allows a test sequence to be stopped while test leads are changed or other operations performed. A 13-character user programmable message will be displayed on the screen when in PAUSE mode and the test will continue when the [START] button is pressed or START initiated via remote I/O.
Page 66
PA Programming – continued from last page. Press DOWN arrow [⇓] Use Function Key to move cursor to next character STEP SETTING Use Numeric Keys then [ENTER] Test Step Test Mode Enter the Message: 0 - 13 alpha-numeric characters NEXT CHAR. Message Under Test PRESS...
Page 67
2.4.5 Programming a DC Hipot Test In a DC hipot test, high DC voltage is applied between the DUT’s operating circuits and chassis ground to determine if/when a breakdown will occur in the insulation of the DUT. The current is measured between the DUT insulation and ground. A high limit is required in a DC test.
Page 68
DC Hipot Programming – continued from last page. Press DOWN arrow [⇓] Use Numeric Keys then [ENTER] STEP SETTING Test Step Enter the Test Voltage in kilo-volts (kV) Test Mode Voltage 0.05 - 6.0 kV (50 - 6000V) High Limit 0.500mA Low Limit Dwell Time...
Page 69
DC Hipot Programming – continued from last page. Press DOWN arrow [⇓] Use Numeric Keys then [ENTER] STEP SETTING Test Step Enter the Dwell Time in seconds (s) Test Mode Voltage 2.400 kV 0 - 999 seconds; 0 = OFF High Limit 0.500mA Low Limit...
Page 70
DC Hipot Programming – continued from last page. Press DOWN arrow [⇓] Use Numeric Keys then [ENTER] STEP SETTING Test Step Enter the Test Time in seconds (s) Test Mode Voltage 2.400 kV 0, 0.3 - 999 seconds; 0 = Continuous High Limit 0.500mA Low Limit...
Page 71
DC Hipot Programming – continued from last page. Scanner #1 is the internal 6000-07 Hipot/Leakage Current Scanner. In DC mode, only Channels 1 and 3 are available on Scanner #1. CH1 can be set to H or X. CH3 can be set to H, L or X.
Page 72
2.4.6 Programming an Insulation Resistance (IR) Test In an IR test, high DC voltage is applied to the DUT to determine strength of the insulation of the DUT. The resistance is measured between the DUT insulation and ground. A low limit is required in an IR test. The low limit is the minimum allowable value for the test to be considered a PASS.
Page 73
IR Programming – continued from last page. Press DOWN arrow [⇓] Use Numeric Keys then [ENTER] STEP SETTING Test Step Enter the Test Voltage in kilo-volts (kV) Test Mode Voltage 0.05 - 1 kV (50 - 1000V) 1.0M Ω Low Limit High Limit Test Time 3.0s...
Page 74
IR Programming – continued from last page. Press DOWN arrow [⇓] Use Numeric Keys then [ENTER] STEP SETTING Test Step Enter the Test Time in seconds (s) Test Mode Voltage 1.000kV 0, 0.3 - 999 seconds; 0 = Continuous Low Limit 10,000MΩ...
Page 75
IR Programming – continued from last page. Scanner #1 is the internal 6000-07 Hipot/Leakage Current Scanner. In IR mode, only Channels 1 and 3 are available on Scanner #1. CH1 can be set to H or X. CH3 can be set to H, L or X.
Page 76
2.4.7 Programming a Leakage Current (LC) Test In an LC test, high voltage is applied to the DUT and the resistance is measured between the DUT insulation and ground. A high limit is required in an LC test. The high limit is the maximum allowable value for the test to be considered a PASS.
Page 77
LC Programming – continued from last page Press DOWN arrow [⇓] UseFunction Key STEP SETTING UL544NP Test Step Select Equivalent Circuit Test Mode For Human Body Impedance Model UL544P Device Line Input NORMAL UL1563 GB Switch OPEN Meter UL2601 High Limit 6.000mA Low Limit UL1950...
Page 78
LC Programming – continued from last page Press DOWN arrow [⇓] Use Function Key STEP SETTING Line to GND Earth L - G Test Step Test Mode Line to Patient Probe #2 Enclosure, Patient Device UL2601 L - P2 Line Input NORMAL OPEN Pat.
Page 79
LC Programming – continued from last page Press DOWN arrow [⇓] Use Function Key STEP SETTING Setup Power Measurement SETUP Test Step Test Mode Device UL2601 Line Input NORMAL GB Switch OPEN Meter L - G 10.00mA High Limit Low Limit VOLTAGE Power 3.0s...
Page 80
LC Programming – continued from last page. Press DOWN arrow [⇓] Use Numeric Keys then [ENTER] STEP SETTING Test Step Enter the Power Low Limit or Target Frequency Test Mode [VOLTAGE] Power Low: 0-300V, 0 = OFF Device UL2601 POWER SETUP [CURRENT] Power Low: 0-20A, 0 = OFF [VA]...
Page 81
LC Programming – continued from last page. Scanner #1 is the internal 6000-07 Hipot/Leakage Current Scanner. In LC mode, Channels 1 and 3 are available on Scanner #1. CH1 is used to provide the voltage to the DUT to power it on. CH3 can be set to L or X. Use Numeric Keys then [ENTER] STEP SETTING Use the Numeric Key that corresponds to the Channel #...
Page 82
2.4.7.1 Measurement Equipment A critical specification for IEC60601-1 standard safety testing is the actual measurement device used in the testing. IEC60601-1 requires the use of a true RMS multi-meter with a very specific load. The load simulates the impedance of the human body. Figure 2-8 illustrates the IEC60601-1 required test load.
Page 83
G6100 Plus Leakage Current Tests The Guardian 6100 Plus Medical Safety Analyzer measures Earth Line Leakage, Enclosure Leakage, Patient Line Leakage and Patient Auxiliary Leakage Current. The instrument simulates the human body models listed in Table 2-8. The test modes and other programming parameters are also listed in Table 2-8.
Page 84
Leakage Current – Connection Using In-Rush Protection When using the Guardian 6100 Plus with a device that may have high in-rush current the isolation transformer should be connected using In-Rush Protectors as shown in Figure 2.11. The In-Rush Protector (QuadTech part number 800184) is a standard accessory included with the 6100 Plus instrument.
Page 85
The Guardian 6100 Plus can perform this test by just plugging the medical device into the G30 corded product adapter. The tests can be performed with any of 4 different mains conditions.
Page 86
Earth Leakage: Connection to DUT Figure 2-12b illustrates the connection of the Guardian 6100 Plus for an Earth Leakage Current test. Connect the G32 Isolation Transformer to the 6000-07 LINE INPUT terminals (L-IN and N-IN). Connect the black leads of the G30 Corded Product Adapter to the black (low) SENSE and DRIVE “Optional Rear Output Terminals”...
Page 87
2.4.7.3 Enclosure (Touch/Chassis) Leakage Current flowing from the enclosure or other parts, excluding patient connections, back to ground through an impedance which simulates the impedance of the human body. This is the leakage current that would flow through an individual’s body should they “touch” that part of the medical device.
Page 88
Enclosure (Touch/Chassis) Leakage: Connection to DUT Figure 2-13b illustrates the connection of the Guardian 6100 Plus for an Enclosure Leakage Current test. Connect the G32 Isolation Transformer to the 6000-07 LINE INPUT terminals (L-IN and N-IN). Connect the black leads of the G30 Corded Product Adapter to the black (low) SENSE and DRIVE “Optional Rear Output Terminals”...
Page 89
2.4.7.4 Patient Leakage – Applied Part to Ground There are two patient connections P1 and P2 on the back of the Guardian 6100 Plus. Patient Leakage can be measured from patient connection P2 to earth ground. This test is performed identical to enclosure leakage current.
Page 90
Patient Leakage – Applied Part to Ground: Connection to DUT Figure 2-14b illustrates the connection of the Guardian 6100 Plus, G30, G32 and device under test for the Patient Leakage – Applied Part to Ground test. The external connections are the same as other leakage tests except for the patient connection, P2.
Page 91
2.4.7.5 Patient Auxiliary Leakage The IEC 60601-1 measuring circuit for the patient auxiliary current test is illustrated in Figure 2-15a. Patient Auxiliary Current can also be measured from patient connection P1 to all other patient connections P2. The tests can be performed with any of 4 different mains conditions. These conditions are: Normal Mains (S5 DOWN position) Reverse Mains (Hot and Neutral reversed) (S5 UP position, as shown)
Page 92
Patient Auxiliary Leakage: Connection to DUT Figure 2-15b illustrates the connections between the Guardian 6100 Plus, G30 corded product adapter, G32 isolation transformer and device under test for a Patient Auxiliary Leakage current test. QuadTech Device UnderTest Plug into G30...
Page 93
2.4.8 Programming a Multi-Step Test A total of 100 tests can be programmed and each test can consist of 1 to 50 steps in sequence. Each step may be programmed for any available function (GB, AC, DC, IR, LC, PA or OSC) with programmed test conditions independent from the other step. Contained in this paragraph are the text instructions to program the example 7-step test as shown in Table 2-5 in ¶...
Page 94
Press the function key [AC] to select mode = AC Hipot Press DOWN arrow [⇓] Press [1][.][5][0][0][ENTER] to set voltage = 1.500kV (1500V) Press DOWN arrow [⇓] Press [0][.][5][0][0][ENTER] to set the high current limit = 0.500mA Press DOWN arrow [⇓] Press [0][ENTER] to set low current limit = OFF Press DOWN arrow [⇓] Press [0][ENTER] to set arc current limit = OFF...
Page 95
Press the function key [3 -230 kHz] to set arc filter frequency = 3 – 230kHz Press DOWN arrow [⇓] Press [3][.][0][ENTER] to set test time = 3.0s Press DOWN arrow [⇓] Press [3][.][0][ENTER] to set ramp time = 3.0s Press DOWN arrow [⇓] Display reads: “11.
Page 96
Press the function key [VOLTAGE] to set power = Voltage mode Press DOWN arrow [⇓] Press [3][0][0][ENTER] to set high power limit = 300V Press DOWN arrow [⇓] Press [0][ENTER] to set low power limit = OFF Press the function key [EXIT] to exit power setup Press DOWN arrow [⇓] Press [3][.][0][ENTER] to set test time = 3.0s Press DOWN arrow [⇓]...
Page 97
MEMORY There are 100 locations in memory and up to 50 test steps may be stored in any one location for a total memory block of 500. There are 10 memory locations listed per page on the display. [NEXT PAGE] will move the backlit box to the top of the next page in memory.
Page 98
To access the STORE function in MAIN MENU, press the function key [MEMORY]. Use the [↑] and [↓] arrows to move the backlit box to the desired memory location (the number in parenthesis will be 0 indicating no steps and the backlit box will be empty) _____________.
Page 99
2.5.4 Clearing Memory All stored test conditions in instrument memory (100 setups) may be cleared. Caution: all preset and option parameters are cleared as well as all memory locations. All parameters will be reset to default values. To clear setup memory: With the instrument in MAIN MENU, Press the numerical key [4] that corresponds to CALIBRATION.
Page 100
Table 2-9: Preset Parameters Parameter Range Default Pass Hold 0.2 – 99.9s 0.5s Step Hold KEY, 0.1 – 99.9s 0.3s AC Frequency 50 – 600Hz 60Hz GB Frequency 50/60Hz 60Hz IEC-601 ON/OFF GB Voltage 6.0 – 15.0V 15.0V Auto Range ON/OFF Software AGC ON/OFF...
Page 101
PRESET SETUP Pass Hold 0.5 sec 0.3 sec Step Hold AC Freq. 64 Hz GB Freq. 60 Hz IEC - 601 GB Voltage 15.0 V Auto Range Soft AGC Part No. Lot No. Serial No. Start W ait KEY, 0.1 - 99.9s Remote Lock Offset...
Page 102
PRESET SETUP Pass Hold 0.5 sec Step Hold 0.3 sec AC Freq. 64 Hz 50 Hz 60 Hz GB Freq. IEC - 601 GB Voltage 15.0 V Auto Range 60 Hz Soft AGC Part No. Lot No. Serial No. Start W ait 50/60 Hz Remote Lock...
Page 103
PRESET SETUP Pass Hold 0.5 sec Step Hold 0.3 sec AC Freq. 64 Hz GB Freq. 60 Hz IEC - 601 15.0 V GB Voltage Auto Range Soft AGC Part No. Lot No. Serial No. Start W ait 6.0 - 15.0V Remote Lock Offset...
Page 104
2.6.9 Part No. The PART NUMBER function allows the operator to assign a part number to the device under test. To access Part No., press [PRESET] and then the [↓] or [↑] arrow on the keypad so the backlit box is to the right of Part No. Use the numerical keys to select first character then use [NEXT CHAR] to move the underscore cursor to the next character.
Page 105
2.6.11 Serial No. The SERIAL NUMBER function allows the operator to assign a serial number to the device under test. To access Serial No., press [PRESET] and then the [↓] or [↑] arrow on the keypad so the backlit box is to the right of Serial No. Use the numerical keys to select first character then use [NEXT CHAR] to move the underscore cursor to the next character.
Page 106
2.6.12 Start Wait The Start Wait time is a delay time – after connection to the device is made and before current is applied to begin the ground bond test. This feature is normally used in a probing application. For example: If a 2 second start wait time is programmed, once the continuity between high and low GB connections is made, the G6100 Plus unit waits the 2sec programmed start time then applies current to the DUT.
Page 107
Menu. 2.7.1 Offset The Guardian 6100 Plus provides automatic offset for lead or fixture effects. During the offset process a correction is made (subtracted out) as the result of lead leakage current and stored in instrument memory to be applied to ongoing measurements. For maximum measurement accuracy it is recommended that offset be performed after power-up, any time the test parameters are changed and any time the test leads or fixture are changed.
Page 108
QuadTech Guardian 6100 Plus STOP Twin Electrical Safety Analyzer MENU PRINT LOCAL Por t s DANGER HIGH VOLTAGE START CAUTION! Max: 5kVAC 6kVDC OUTPUT PASS FAIL Drive Sense Sense Drive ENTER UPDATE DRIVE+ and DRIVE- NOTE: Output Terminals Make the connection to the DRIVE+ and DRIVE- terminals with the spade lug behind the nut.
Page 109
SHORT the Product Alligator Drive- & Drive+ Adapter Terminals: Cable Place banana plug in ground blade socket QuadTech 125V DO NOT TOUCH W HILE UNDER TEST MAX. VOLTAGE 5kVAC 6kVDC (Monitor) Figure 2-24: Offset Short using G15 and G30 Corded Product Adapter Offset –...
Page 110
TEST - M003 - EXAMPLE MODE SOURCE LIMIT RES. OFFSET 100.0m Ω 30.0A Open GET Cs 1.500kV 0.500mA PAGE UP CHANGE LEAD 2.400kV 0.500mA PAGE DOWN 10.0G Ω 1.000kV 0.300kV 10.00mA SCANNER - 1 1 2 3 4 5 6 7 8 Remote Lock Offset...
Page 111
Offset – continued The following formulas apply to the offset function: For DC offset: Display current = (current read) - (offset current) For AC offset: If Measured Value < Value set in ¶ 2.3.1.9: − Display current = current read offset current If Measured Value >...
Page 112
The leads consist of a white plug for connection to HV OUTPUT and a black banana plug with retaining bracket for connection to the Drive- terminal. Figure 2-25 illustrates the connection of the S-02 Cable Lead Set to the G6100 Plus instrument. QuadTech Guardian 6100 Plus STOP Twin Electrical Safety Analyzer...
Page 113
T20A / 250V Corded Connected Product Banana to to exposed Adapter Alligator metal on Cable QuadTech Plug into G30 125V DO NOT TOUCH W HILE UNDER TEST MAX. VOLTAGE 5kVAC 6kVDC (Monitor) Figure 2-26: G30 Connection to G6100 Plus for GB & Hipot...
Page 114
QuadTech Guardian 6100 Plus STOP Twin Electrical Safety Analyzer MENU PRINT LOCAL Por t s White Custom Banana DANGER HIGH VOLTAGE Plug START CAUTION! Max: 5kVAC OUTPUT 6kVDC OUTPUT PASS FAIL Drive Sense Sense Drive ENTER UPDATE Black Spade to Drive-...
Page 115
The red banana plug and spade lug connect to the (+) side of the out put terminals and product. Make the connection to the Drive + and Drive – terminals with the spade lug positioned behind the nut. QuadTech Guardian 6100 Plus STOP Twin Electrical Safety Analyzer...
Page 116
Connect the white plug to HV OUTPUT. The black banana plug with retaining bracket for connection to the Drive- terminal is not needed when using the G15 with the S02. QuadTech Guardian 6100 Plus STOP Twin Electrical Safety Analyzer...
Page 117
Connect the white plug to HV OUTPUT, the black banana plug with retaining bracket to the Sense- terminal, and the black spade lug to the Drive- terminal. QuadTech Guardian 6100 Plus STOP Twin Electrical Safety Analyzer...
Page 118
* Netherlands * Austria * Switzerland * Italy Refer to Figure 2-31 for connection of the G16 International Power Strip to the Guardian 6100 Plus instrument output terminals. QuadTech Guardian 6100 Plus STOP Twin Electrical Safety Analyzer MENU PRINT LOCAL...
Page 119
G45 INPUT terminals. The red wire between Drive +/Drive+. The black wire between Drive-/Drive-. The blue spades to Sense+ and Sense- terminals. Connect the Banana/Spade-to-Alligator lead set between the G45 OUTPUT terminals and the DUT. QuadTech Guardian 6100 Plus STOP Twin Electrical Safety Analyzer MENU PRINT...
Page 120
2.7.4 Measurement Procedure Before a measurement is made, verify the following: • Guardian 6100 Plus unit power is ON • 15 minute warm-up • Test parameters programmed • Offset function completed • OSC – Get Cs completed • Device Under Test (DUT) connected Perform a test at power-up conditions (test conditions at which the instrument was last powered down), or recall one of 100 possible stored setups.
Page 122
Figures 3-2 and 3-3 illustrate possible remote control connections. Use extreme care when using a remote control connection as the High Voltage Output is being turned ON and OFF with with an external signal. UNDER TEST START RESET PASS INTER FAIL LOCK RESET...
Page 123
PASS CONDITION FAIL CONDITION CLOSE START OPEN CLOSE UNDER TEST OPEN 45ms 45ms CLOSE PASS OPEN 300us CLOSE FAIL OPEN 300us CLOSE STOP OPEN CLOSE INTERLOCK OPEN Figure 3-4: G6100 Plus Timing Diagram Interface 150799 Rev A3 Page 123 of 177 www.valuetronics.com...
Page 124
RS232 Interface 3.2.1 PIN Configuration An RS232 interface is standard equipment on the Guardian 6100 Plus instrument. Connection is through the black 9-PIN connector labeled ‘RS232’ on the rear panel of the Guardian 6100 Plus unit. Figure 3-5 illustrates the RS232 PIN configuration.
Page 125
3.2.2 RS232 Specifications Data bits: 8 (or 7 data bits and 1 parity bit) Stop bits: Start dits: Parity: None/Odd/Even Baud Rate: 300/600/1200/2400/4800/9600/19200 bps Flow Control: None/Software EOS: CR + LF Selecting Baud Rate Setting the baud rate is done in the [OPTION] submenu, ¶ 2.3.2.1. Use RS-232 to set the baud rate, parity and flow control for the RS-232 interface.
Page 126
3.2.4 Sample QuickBasic Program REM Guardian 6100 Plus RS232 example program REM Created using Quick Basic REM open serial port as device 1 OPEN "COM1:9600,N,8,1,RS," FOR RANDOM AS #1 PRINT "This program will show the configurable commands for the G6100 Plus"...
Page 127
REM Program test time for 5 sec PRINT #1, "SOURce:SAFEty:STEP2:GB:TIME 2"; CHR$(13); CHR$(10) REM Program high resistance limit 100mOHM PRINT #1, "SOURce:SAFEty:STEP2:GB:LIMit:HIGH 0.1000"; CHR$(13); CHR$(10) REM Program TWIN PORT off PRINT #1, "SOURce:SAFEty:STEP2:GB:TPORT 0"; CHR$(13); CHR$(10) REM Program STEP3 for AC HIPOT REM Program AC test voltage for 1000 V PRINT #1, "SOURce:SAFEty:STEP3:AC:LEVel 1000";...
Page 128
REM Program ramp time for 2 seconds PRINT #1, "SOURce:SAFEty:STEP4:DC:TIME:RAMP 2.0"; CHR$(13); CHR$(10) REM Program dwell time for 1 seconds PRINT #1, "SOURce:SAFEty:STEP4:DC:TIME:DWELL 1.0"; CHR$(13); CHR$(10) REM Program channel 3 high and channel 1 off PRINT #1, "SOURce:SAFEty:STEP4:DC:CHAN:HIGH (@1(3))"; CHR$(13); CHR$(10) PRINT #1, "SOURce:SAFEty:STEP4:DC:CHAN:LOW (@1(0))";...
Page 129
REM Program STEP 7 for UL1950 REM SET DEVICE FOR UL1950 PRINT #1, "SOURce:SAFEty:STEP7:LC:DEVICE UL1950"; CHR$(13); CHR$(10) REM Program LINE FOR Reverse PRINT #1, "SOURce:SAFEty:STEP7:LC:LINE REV"; CHR$(13); CHR$(10) REM Program METER L - G PRINT #1, "SOURce:SAFEty:STEP7:LC:METER L,G"; CHR$(13); CHR$(10) REM Program GROUND SWITCH Closed PRINT #1, "SOURce:SAFEty:STEP7:LC:GSWITCH 1";...
Page 130
REM Program POWER TO VA LOW LIMIT 1000 PRINT #1, "SOURce:SAFEty:STEP8:LC:POWER:VA:LIMIT:LOW 1000"; CHR$(13); CHR$(10) REM PROGRAM PRESET REM PROGRAM PASS HOLD TO 0.5 SEC PRINT #1, "SOUR:SAFE:PRESET:time:pass 0.5"; CHR$(13); CHR$(10) REM PROGRAM STEP HOLD TO 1 SEC PRINT #1, "SOUR:SAFE:PRESET:time:STEP 1"; CHR$(13); CHR$(10) REM PROGRAM AC FREQ TO 100 PRINT #1, "SOUR:SAFE:PRESET:AC:FREQ 100";...
Page 131
IEEE-488 Interface 3.3.1 PIN Configuration An optional IEEE-488 interface is available for the Guardian 6100 Plus instrument. Connection through a blue 24-PIN connector (labeled GPIB) on the rear panel. This interface can be used to connect a system containing a number of instruments and a controller in which each meets IEEE Standard 488.2 (Standard Digital Interface for...
Page 133
Table 3-1: IEEE-488 Interface Pin Designations Signal Function Name Number Low State: “Data is Available” and valid on DI01 through DI08 NRFD Low State: At least one Listener on the bus is “Not Ready For Data” NDAC Low State: At least one Listener on the bus is “Not Accepting Data”...
Page 134
3.3.2 IEEE-488 Interface Codes and Messages The IEEE-488 (GPIB) address is defined in the [OPTION] submenu by selecting [MENU] [2] and [GPIB]. Refer to paragraph 2.3.2.2. The G6100 Plus unit is in a remote control status when the REMOTE LED is ON. To switch to Local from Remote press the [LOCAL] function key, disabled by LLO message.
Page 135
3.3.3 IEEE-488 Interface Commands The interface function is controlled by ASCII commands which include: {[command + parameter] ; [command + parameter] + ending code} The length of the string is 128 characters. It is not necessary to input any sign or space between the command and parameter.
Page 136
3.3.3.2 IEEE-488 Register Configuration Figure 3-7 illustrates the configuration of the Summary (Status), Event and Enable Registers for the Guardian 6100 Plus IEEE-488 Interface. Error/Event Queue Enable Summary Register Register 0 Not Used Has Result Error/Event Queue Output Not Used...
Page 137
Table 3-6: IEEE-488 Commands Command Name Function Output Format *CLS Clear Status Clear standard event status register. Clear status bit group register except for bit 4 (MAV) *ESE Event Status Enable Enable standard event status 0 – 255 register value. *ESE? Event Status Enable Query standard event status of...
Page 138
3.3.4 IEEE-488 & SCPI Command Summary The SCPI (Standard Commands for Programmable Instruments) parameter syntax format is listed on the following pages. To illustrate the order of precedence, the entire command list is scrolled through. The dual arrow symbol “<>” denotes the defined parameter is a standard SCPI command.
Page 139
:FREE :STEP? Command: MEMory : FREE : STEP? This command queries the next free step in main memory. :PRESet? Command: :MEMory : FREE : PRESet? This command queries the next free preset number in main memory. :SOURce :SAFEty :STARt Command: :SOURce : SAFEty : STARt This command starts the test.
Page 140
:RESult :ALL :JUDGment? Command: :SOURce : SAFEty : RESult : ALL : JUDGment? This command queries all the judgment results. The return format is: PASS USER STOP CAN NOT TEST TESTING STOP HEX and DEC values: GB MODE AC MODE DC MODE IR MODE OVER...
Page 141
:AREPort ? (RS232 Interface only) Command: :SOURce : SAFEty : RESult : AREPort? This command queries if the Automatic Reporting of test results is ON or OFF. The return format is a “1” or a “0”. :SNUMber? Command: :SOURce : SAFEty : SNUMber? This command queries the step number being set in memory.
Page 142
:ARC <numeric value> Command: :SOURce : SAFEty : STEP <n> : AC : LIMit : ARC This command sets the arc current limit for the selected step. <numeric value> is 0 to 0.030 and the unit is amps. 0 = OFF :ARC? Command: :SOURce : SAFEty : STEP <n>...
Page 143
:LOW <channel list> Command: :SOURce : SAFEty : STEP <n> : AC : CHAN : LOW (@ Box# (<channel list>)) This command sets the low scan channels for the selected step. :LOW? Command: :SOURce : SAFEty : STEP <n> : AC : CHAN : LOW? This command queries the low scan channels for the selected step.
Page 144
:ARC:FILTER <numeric value> Command: :SOURce : SAFEty : STEP <n> : DC : LIMit : ARC FILTER This command sets the arc filter for the selected step. <numeric value> is 23e3, 50e3, 100e3 or 230e3 and the unit is Hz. :ARC FILTER? Command: :SOURce : SAFEty : STEP <n>...
Page 145
:LOW <channel list> Command: :SOURce : SAFEty : STEP <n> : DC : CHAN : LOW (@Box#(<channel list>)) This command sets the low scan channels for the selected step. :LOW? Command: :SOURce : SAFEty : STEP <n> : DC : CHAN : LOW? This command queries the low scan channels for the selected step.
Page 146
:TEST <numeric value> Command: :SOURce : SAFEty : STEP <n> : IR : TIME : TEST This command sets the test time for the selected step. <numeric value> is 0, 0.3 to 999 and the unit is seconds. 0 = Continuous :TEST? Command: :SOURce : SAFEty : STEP <n>...
Page 147
:LIMit:LOW <numeric value> Command: :SOURce:SAFEty:STEP<n>:GB:LIMit:LOW <numeric value> This command sets the low resistance limit for the selected step. <numeric value> is 0 to 100 and the unit is milliohms. :LIMit:LOW? Command: :SOURce:SAFEty:STEP<n>:GB:LIMit:LOW? Return Value: +1.000000E-02 This command queries the low resistance limit for the selected step. :TIME :TEST <numeric value>...
Page 148
:DEVice Command: :SOURce:SAFEty:STEP<n>:LC:DEVice This command sets the circuit model for the selected step. UL1950 | UL 1563 | UL544NP | UL544P | UL2601 :DEVice? Command: :SOURce:SAFEty:STEP<n>:LC:DEVice? This command queries the circuit model for the selected step. :LINE Command: :SOURce:SAFEty:STEP<n>:LC:LINE This command sets the line input for the selected step. NORmal | REVerse | SFNormal | SFReverse :LINE? Command: :SOURce:SAFEty:STEP<n>:LC:LINE?
Page 149
:LIMit:LOW <numeric value> Command: :SOURce:SAFEty:STEP<n>:LC:LIMit:LOW <numeric value> This command sets the low current limit for the selected step. <numeric value> is 0 to 10 and the unit is milliamps. :LIMit:LOW? Command: :SOURce:SAFEty:STEP<n>:LC:LIMit:LOW? Return Value: +1.000000E-02 This command queries the low current limit for the selected step. :TIME [:TEST]<numeric value>...
Page 150
:CURRent Command: :SOURce:SAFEty:STEP <n>:LC:POWer:CURRent This command sets the power measurement to current for the selected step. :LIMit[:HIGH] <numeric value> Command: :SOURce:SAFEty:STEP<n>:LC:POWer:CURRent[:LIMit][:HIGH]<nv> This command sets the high power limit for the selected step. <numeric value> is 0 to 20 and the unit is amps :LIMit[:HIGH[? Command: :SOURce:SAFEty:STEP<n>:LC: POWer:CURRent[:LIMit][:HIGH]? This command queries the high power limit for the selected step...
Page 151
:SIMUlation Command: :SOURce:SAFEty:STEP <n>:LC:POWer:SIMUlation This command sets the power measurement to simulation for the selected step. :SIMUlation:TVOLtage <numeric value> Command: :SOURce:SAFEty:STEP<n>:LC:POWer:SIMUlation:TVOLtage <nv> This command sets the target voltage for the selected step. <numeric value> is 80 to 300 and the unit is volts :SIMUlation:TVOLtage? Command: :SOURce:SAFEty:STEP<n>:LC: POWer: SIMUlation:TVOLtage? This command queries the target voltage for the selected step...
Page 152
:FREQuency <numerical value> Command: :SOURce : SAFEty : PRESet : GB : FREQuency This command sets the test frequency for GB Test. <numeric value> is a value between 50 or 60 and the unit is Hz. :FREQuency? Command: :SOURce : SAFEty : PRESet : GB : FREQuency? This command queries the test frequency for the GB Test.
Page 153
3.3.5 Sample QuickBASIC Program Refer to paragraph 3.2.4 for QuickBasic RS-232 program example. Printer Interface An optional Printer Interface (G38) is available for the Guardian 6100 Plus instrument. The Printer interface takes the place of the IEEE-488 interface and is factory installed when the unit is ordered.
Page 154
Scanner Interface 3.5.1 Scanner Accessories The Scanner Interface is a standard accessory on the Guardian 6100 Plus instrument. Connection is through the black 25-PIN connector labeled ‘SCAN’ on the rear panel of the G6100 Plus instrument. The G6100 Plus unit has a standard internal 6000-07 scanner and may be used with an external scan unit for multi-point grounding and hipot tests.
Page 155
The rear panel outptut of the G6100 Plus has a Floating Ground connection (“Ch-3” when programming internal scanner). Refer to p52, Figure 2-7. Before connecting the scanner to the Guardian 6100 Plus or connecting devices for test, press the [STOP] key and make sure the red DANGER light is OFF.
Page 156
WARNING THE REAR PANEL GROUND LUGS ON ALL INSTRUMENTS (Guardian 6100 Plus and Scanners) MUST BE INTERCONNECTED GB FLOATING SENSE DRIVE HV OUTPUT N-OUT (CH-1) L-OUT (CH-1) PATIENT PROBE LINE INPUT INPUT FUSE N-IN L-IN Connected Banana to to exposed...
Page 157
Use Function Key STEP SETTING SETUP Test Step Enter setup menu for scanner(s) Test Mode Device UL2601 Line Input NORMAL GB Switch OPEN L - G Meter 10.00mA High Limit Low Limit Power VOLTAGE Test Time 3.0s CHNL (H-L) PRESS FUNCTION KEY Remote Lock Offset...
Page 158
3.5.4 Programming an External Scanner NOTE: Connection to External Scanners (5000-01 to -04) should be made via the front panel outptut connectors of the G6100 Plus instrument. The rear panel output connectors of the G6100 Plus are not identical to the front panel output connectors.
Page 159
Program the Scanner Channels Continuing the example, there are 4 DUTS connected to each of the 5000-01 external scanners as illustrated in Figure 3-13. The internal 6000-07 scanner is not used in this example. Program an OSC and an AC test and set CH1, 3, 5 & 7 to “L” and CH2, 4 6 & 8 to “H”...
Page 160
Press the function key [3 -230 kHz] to set arc filter frequency = 3 – 230kHz Press DOWN arrow [⇓] Press [3][.][0][ENTER] to set test time = 3.0s Press DOWN arrow [⇓] Press [0][ENTER] to set ramp time = OFF Press DOWN arrow [⇓] Display reads: “10.
Page 161
External Scanner Connection – Multi-Point Hipot Figure 3-13 illustrates a multi-point Hipot/IR setup. The Guardian 6100 Plus with is connected in a chain with two external Guardian 5000-01 scanners. SCANNERS #4 - 8 REAR FRONT QuadTech 5000-01 Scanner PANEL PANEL...
Page 162
External Scanner Connection – Multi-Point Ground Bond Figure 3-14 illustrates the Guardian 6100 Plus connected to an external Guardian 5000- 02 scanner for multi-channel Ground Bond testing. The external scanner must be connected to the FRONT panel output connectors of the G6100 Plus. The 6000-07 Hipot/Line Leakage internal scanner is shown installed but is not used in this GB application.
Page 163
Troubleshooting Listed herein are some common errors and suggested methods to resolve the problem. High Limit Failures: 1. Check DUT ensure that it is not faulty 2. Remove DUT and perform test with cables only. If the test passes, ensure offset LED is on (if this function is being used), if offset LED is off, possible cause of failure is the additional leakage from offset being off.
Page 164
No Display: Adjust contrast by pressing [MENU], [1] and the [↓] or [↑] arrow on the keypad so the backlit box is to the right of Contrast. Use the function keys [UP] or [DOWN] to increase or decrease the contrast. The display Contrast is adjustable from 1 –...
Page 165
Shipments sent collect can not be accepted. Calibration Calibration of the Guardian 6100 Plus is recommended on an annual basis. If the unit is returned to QuadTech for factory calibration refer to paragraph 4.2 for instructions. Using the calibration procedure below the instrument can be calibrated by a qualified service person if traceable calibration equipment and standards are available.
Page 166
Table 4-1 : Calibration Equipment Equipment Requirements AC/DC High Voltage Voltmeter Measure Range: 0 to 4kV, 0.1% accuracy AC/DC Current Meter Measure Range: 0 to 10mA, 0.15% accuracy IR Standards Box Tool # 01-1395 or equivalent 250V 1GΩ Resistance Standard 500V 100MΩ...
Page 167
4.3.1.1 AC Voltage Calibration Connect the OUTPUT terminal of the Guardian 6100 Plus to the input terminal of the AC/DC high voltage meter. Connect the RTN/LOW terminal of the Guardian to the GND terminal of the voltmeter. Set the voltmeter to AC and 200V range.
Page 168
4.3.1.4 AC Current Calibration Connect the OUTPUT terminal of the Guardian 6100 Plus to a resistance box or resistance standard. Connect an AC/DC current meter in series between the resistance (box) and RTN/LOW terminal. Set the meter to AC current. Table 4-3 lists the loads for the current calibration steps.
Page 169
To continue calibration from step 6 proceed as follows: Press [UP] to go to CAL step 7. (ACA 3mA Offset 0.12mA) Press [STOP] Press [START] Using keypad enter the reading from the current meter followed by the [ENTER]. Press [STOP] to accept reading. Press [UP] to go to CAL step 8.
Page 170
4.3.1.5 DC Current Calibration Make sure to change the meter’s setting to DC current prior to STEP 11. Ensure the Current meter is set to the lowest range possible, for each step. To continue calibration from step 10 proceed as follows: Press [UP] to go to CAL step 11.
Page 171
Table 4-4 lists the loads for the Ground Bond tests. Connect the Drive & Sense terminals of the Guardian 6100 Plus to the 1mΩ r esistance standard as shown in Figure 4-1. Connect the AC voltage meter to the Sense + and Sense – terminals of the Guardian 6100 Plus.
Page 172
2 wire 53.71 2 wire 102.45 4 wire 10.56 4 wire 50.40 4 wire 99.16 Source Sense Source Sense Source Sense QuadTech Guardian 6100 Plus STOP Twin Electrical Safety Analyzer MENU PRINT LOCAL Por t s DANGER HIGH VOLTAGE START...
Page 173
4.3.1.7 Ground Bond Voltage Calibration: With 100mΩ load attached: Press [UP] to go to CAL step 17. (GBV 8V Offset 0.3V) Press [STOP] Press [START] Using keypad enter the voltage reading from the voltmeter followed by [ENTER]. Press [STOP] to accept reading. Press [UP] to go to CAL step 18.
Page 174
Continue with steps 22 -25 by attaching the dummy load box outlined in Table 4-5 between the High voltage output and the return. Press [UP] to go to CAL step 22. (IRR 250V Range1 1GΩ) Press [STOP] Press [START] Enter the value of the resistor standard followed by [ENTER] Press [STOP] Press [UP] to go to CAL step 23.
Page 175
4.3.1.10 LC Meter Calibration Connect an AC voltage source (Freq = 60Hz) between the N-IN terminal on the 6000-07 scanner and DRIVE- on the GB Floating terminals. Press [UP] to go to CAL step 1. (LC 1V Offset 0.08V) Press [STOP] Press [START] Using keypad enter the reading from the voltmeter followed by [ENTER].
Page 176
4.3.1.11 LC Power Meter Calibration LC Voltage 300V Offset and Full Connect an AC voltage source (Freq = 60Hz) between the L-IN terminal and the N-IN terminal on the 6000-07 scanner. Press [UP] to go to CAL step 5. (LCV 300V Offset 20V) Press [STOP] Press [START]...
Page 177
LC Current 10A Offset and Full SAME connection of an AC current source (Freq = 60Hz) between the L-IN terminal and the L-OUT terminal on the 6000-07 scanner. Press [UP] to go to CAL step 9. (LCA 10A Offset 0.8A) Press [STOP] Press [START] Using keypad enter the reading from the ammeter followed by [ENTER].
Need help?
Do you have a question about the Guardian 6100 Plus and is the answer not in the manual?
Questions and answers