DC Hipot Programming – continued from last page.
Press DOWN arrow [⇓]
STEP SETTING
01.
Test Step
02.
Test Mode
03.
Voltage
04.
High Limit
05.
Low Limit
06.
Dwell Time
07.
Arc Limit
08.
Arc Filter
09.
Test Time
10.
Ramp Time
11.
CHNL (H-L)
0.05 - 6.0 kV
Press [2][.][4][0][0][ENTER] to set voltage = 2.400kV
Press DOWN arrow [⇓]
STEP SETTING
01.
Test Step
02.
Test Mode
03.
Voltage
04.
High Limit
05.
Low Limit
06.
Dwell Time
07.
Arc Limit
08.
Arc Filter
09.
Test Time
10.
Ramp Time
11.
CHNL (H-L)
0.0001 - 12 mA
Press [0][.][5][0][0][ENTER] to set high current limit = 0.500mA
Press DOWN arrow [⇓]
STEP SETTING
01.
Test Step
02.
Test Mode
03.
Voltage
04.
High Limit
05.
Low Limit
06.
Dwell Time
07.
Arc Limit
08.
Arc Filter
09.
Test Time
10.
Ramp Time
11.
CHNL (H-L)
0 - 12 mA,
Press [0][ENTER] to set low current limit = OFF
Continued on next page.
Page 68 of 177
www.valuetronics.com
:
5
:
DC
OFF
:
:
0.500mA
:
OFF
:
OFF
:
OFF
:
3-230 kHz
:
3.0s
:
OFF
OFF
:
Remote
Lock
:
5
:
DC
:
2.400 kV
0.500mA
:
:
OFF
:
OFF
:
OFF
:
3-230 kHz
:
3.0s
:
OFF
OFF
:
Remote
Lock
:
5
:
DC
:
2.400kV
:
0.500mA
:
OFF
OFF
:
OFF
:
:
3-230 kHz
:
3.0s
:
OFF
OFF
:
0 = OFF
Remote
Lock
Offset
Error
Offset
Error
Offset
Error
150799 Rev A3
Use Numeric Keys then [ENTER]
Enter the Test Voltage in kilo-volts (kV)
0.05 - 6.0 kV (50 - 6000V)
Use Numeric Keys then [ENTER]
Enter the High Limit in milliamps (mA)
0.0001 - 12 mA
Note:
Setting the High Limit will
determine the measurement range:
<.300mA = Low range
.300 < 3.00mA = Mid range
> 3.00mA = High range
Use Numeric Keys then [ENTER]
Enter the Low Limit in milliamps (mA)
0 - 12 mA, 0 = OFF