REM Program STEP 7 for UL1950
REM SET DEVICE FOR UL1950
PRINT #1, "SOURce:SAFEty:STEP7:LC:DEVICE UL1950"; CHR$(13); CHR$(10)
REM Program LINE FOR Reverse
PRINT #1, "SOURce:SAFEty:STEP7:LC:LINE REV"; CHR$(13); CHR$(10)
REM Program METER L - G
PRINT #1, "SOURce:SAFEty:STEP7:LC:METER L,G"; CHR$(13); CHR$(10)
REM Program GROUND SWITCH Closed
PRINT #1, "SOURce:SAFEty:STEP7:LC:GSWITCH 1"; CHR$(13); CHR$(10)
REM Program HIGH LIMIT TO 10mA
PRINT #1, "SOURce:SAFEty:STEP7:LC:LIMIT:HIGH 10E-3"; CHR$(13); CHR$(10)
REM Program LOW LIMIT TO 1mA
PRINT #1, "SOURce:SAFEty:STEP7:LC:lIMIT:LOW 1E-3"; CHR$(13); CHR$(10)
REM Program TEST TIME TO 3 SEC
PRINT #1, "SOURce:SAFEty:STEP7:LC:TIME:TEST 3"; CHR$(13); CHR$(10)
REM Program POWER TO CURRENT HIGH LIMIT 20
PRINT #1, "SOURce:SAFEty:STEP7:LC:POWER:CURR:LIMIT:HIGH 20"; CHR$(13); CHR$(10)
REM Program POWER TO Current LOW LIMIT 2
PRINT #1, "SOURce:SAFEty:STEP7:LC:POWER:CURR:LIMIT:LOW 2"; CHR$(13); CHR$(10)
REM Program STEP 8 for UL544NP
REM SET DEVICE FOR UL544NP
PRINT #1, "SOURce:SAFEty:STEP8:LC:DEVICE UL544NP"; CHR$(13); CHR$(10)
REM Program LINE FOR Single Fault Normal
PRINT #1, "SOURce:SAFEty:STEP8:LC:LINE SFN"; CHR$(13); CHR$(10)
REM Program METER P1 - P2
PRINT #1, "SOURce:SAFEty:STEP8:LC:METER P,P"; CHR$(13); CHR$(10)
REM Program GROUND SWITCH OPEN
PRINT #1, "SOURce:SAFEty:STEP8:LC:GSWITCH 0"; CHR$(13); CHR$(10)
REM Program HIGH LIMIT TO 100uA
PRINT #1, "SOURce:SAFEty:STEP8:LC:LIMIT:HIGH 100E-6"; CHR$(13); CHR$(10)
REM Program LOW LIMIT TO 1uA
PRINT #1, "SOURce:SAFEty:STEP8:LC:lIMIT:LOW 1E-6"; CHR$(13); CHR$(10)
REM Program TEST TIME TO 3 SEC
PRINT #1, "SOURce:SAFEty:STEP8:LC:TIME:TEST 3"; CHR$(13); CHR$(10)
REM Program POWER TO VA HIGH LIMIT 2000
PRINT #1, "SOURce:SAFEty:STEP8:LC:POWER:VA:LIMIT:HIGH 2000"; CHR$(13); CHR$(10)
Interface
150799 Rev A3
Page 129 of 177