:CHANnel[:LOW]
Command: :SOURce:SAFEty:STEP<n>:LC:CHANnel[:LOW] <channel list>
This command sets the Low (Return) channels for the selected step.
:CHANnel[:LOW]?
Command: :SOURce:SAFEty:STEP<n>:LC:CHANnel[:LOW]?
This command queries the Low (Return) channels for the selected step.
:PRESet
:TIME
:AC
Interface
www.valuetronics.com
:SIMUlation
Command: :SOURce:SAFEty:STEP <n>:LC:POWer:SIMUlation
This command sets the power measurement to simulation for the selected step.
:SIMUlation:TVOLtage <numeric value>
Command: :SOURce:SAFEty:STEP<n>:LC:POWer:SIMUlation:TVOLtage <nv>
This command sets the target voltage for the selected step.
<numeric value> is 80 to 300 and the unit is volts
:SIMUlation:TVOLtage?
Command: :SOURce:SAFEty:STEP<n>:LC: POWer: SIMUlation:TVOLtage?
This command queries the target voltage for the selected step
The return format is a numerical value: 80 ≤ value ≤ 300V
:PASS <numerical value>
Command: :SOURce : SAFEty : PRESet : TIME : PASS
This command sets the time the remote I/O pass relay is closed when result is pass.
<numeric value> is a value between 0.2 and 99.9 and the unit is seconds.
:PASS?
Command: :SOURce : SAFEty : PRESet : TIME : PASS?
This command queries the set time the remote I/O pass relay is to be closed.
:STEP <numerical value> ⏐KEY
Command: :SOURce : SAFEty : PRESet : TIME : STEP
This command sets the interval time between steps.
<numeric value> is a value or character key between 0.1 and 99.9 seconds.
:STEP?
Command: :SOURce : SAFEty : PRESet : TIME : STEP?
This command queries the interval time between tests.
:FREQuency <numerical value>
Command: :SOURce : SAFEty : PRESet : AC : FREQuency
This command sets the test frequency for AC Hipot Test.
<numeric value> is a value between 50 and 600 and the unit is Hz.
:FREQuency?
Command: :SOURce : SAFEty : PRESet : AC : FREQuency?
This command queries the test frequency for the AC Hipot Test.