OSC Programming – continued from last page.
Press DOWN arrow [⇓]
STEP SETTING
01.
Test Step
02.
Test Mode
03.
Open Chk.
04.
Short Chk.
05.
CHNL (H-L)
10% - 100%
Press [5][0][ENTER] to set the Open Check = 50%
Press DOWN arrow [⇓]
STEP SETTING
01.
Test Step
02.
Test Mode
03.
Open Chk.
04.
Short Chk.
05.
CHNL (H-L)
0, 100% - 500%; 0 = OFF
Press [3][0][0][ENTER] to set the Short Check = 300%
Press DOWN arrow [⇓]
Display reads: "05. CHNL (H-L) : OFF"
STEP SETTING
01.
Test Step
02.
Test Mode
03.
Open Chk.
04.
Short Chk.
05.
CHNL (H-L)
PRESS FUNCTION KEY
Press the function key [SETUP] to enter scanner setup menu
Continued on next page.
Operation
www.valuetronics.com
:
2
:
OSC
50%
:
:
300%
:
OFF
Remote
Lock
Offset
:
2
:
OSC
50%
:
300%
:
:
OFF
Remote
Lock
Offset
:
2
:
OSC
50%
:
300%
:
OFF
:
Remote
Lock
Offset
150799 Rev A3
Use Numeric Keys then [ENTER]
Enter the Open comparison value in percent (%)
10% - 100%, default = 50%
Error
Use Numeric Keys then [ENTER]
Enter the Short comparison value in percent (%)
0, 100% - 500%; 0 = OFF, default = 300%
Error
Use Function Key
SETUP
Enter setup menu for scanner(s)
Error