Download Print this page

QuadTech Guardian 6100 Plus Instruction Manual page 79

Electrical safety analyzer

Advertisement

LC Programming – continued from last page
Press DOWN arrow [⇓]
STEP SETTING
01.
Test Step
Test Mode
02.
03.
Device
04.
Line Input
05.
GB Switch
06.
Meter
07.
High Limit
08.
Low Limit
09.
Power
10.
Test Time
11.
CHNL (H-L)
PRESS FUNCTION KEY
Press the function key [SETUP] to setup the power measurement
Press DOWN arrow [⇓]
STEP SETTING
Test Step
01.
02.
Test Mode
03.
Device
1.
Power
2.
Power High
3.
Power Low
10.
Test Time
11.
CHNL (H-L)
PRESS FUNCTION KEY
Press the function key [VOLTAGE] to set power = Voltage mode
Press DOWN arrow [⇓]
STEP SETTING
01.
Test Step
02.
Test Mode
03.
Device
1.
Power
2.
Power High
3.
Power Low
10.
Test Time
11.
CHNL (H - L)
0 - 300V
Press [3][0][0][ENTER] to set high power limit = 300V
Continued on next page.
Operation
www.valuetronics.com
:
7
LC
:
:
UL2601
:
NORMAL
:
OPEN
:
L - G
10.00mA
:
OFF
:
VOLTAGE
:
3.0s
:
OFF
:
Remote
Lock
:
7
LC
:
:
UL2601
POWER SETUP
VOLTAGE
:
OFF
:
OFF
:
3.0s
:
OFF
:
Remote
Lock
:
7
:
LC
:
UL2601
POWER SETUP
VOLTAGE
:
OFF
:
OFF
:
:
3.0s
:
OFF
0=OFF
Remote
Lock
SETUP
Offset
Error
VOLTAGE
CURRENT
VA
SIMULATE
EXIT
Offset
Error
EXIT
Offset
Error
150799 Rev A3
Use Function Key
Setup Power Measurement
Use Function Key
Volts (V)
Current (A)
Volt-Amperes (VA)
Use Source: OFF
[SOURCE] if Use Source: ON
Use Numeric Keys then [ENTER]
Enter the Power High Limit or Target Voltage
[VOLTAGE]
Power High: 0-300V, 0 = OFF
[CURRENT]
Power High: 0-20A, 0 = OFF
[VA]
Power High: 0-4400V, 0 = OFF
[SIMULATE]
Target-V: 80 - 300V
[SOURCE]
Target-V: 80 - 300V
Page 79 of 177

Advertisement