LC Programming – continued from last page.
Press DOWN arrow [⇓]
STEP SETTING
01.
Test Step
02.
Test Mode
03.
Device
1.
Power
2.
Power High
3.
Power Low
10.
Test Time
11.
CHNL (H - L)
0 -300V
Press [0][ENTER] to set low power limit = Off
Press DOWN arrow [⇓]
STEP SETTING
01.
Test Step
02.
Test Mode
03.
Device
04.
Line Input
05.
GB Switch
06.
Meter
07.
High Limit
Low Limit
08.
09.
Power
10.
Test Time
11.
CHNL (H - L)
0, 0.3 - 999s
Press [3][.][0][ENTER] to set test time = 3.0s
Press DOWN arrow [⇓]
Display reads: "11. CHNL (H-L) : OFF"
STEP SETTING
01.
Test Step
02.
Test Mode
03.
Device
04.
Line Input
05.
GB Switch
06.
Meter
07.
High Limit
08.
Low Limit
09.
Power
10.
Test Time
11.
CHNL (H-L)
PRESS FUNCTION KEY
Press the function key [SETUP] to enter scanner setup menu
Continued on next page.
Page 80 of 177
www.valuetronics.com
:
7
:
LC
:
UL2601
POWER SETUP
VOLTAGE
:
300.0V
:
OFF
:
:
3.0s
:
OFF
0=OFF
Remote
Lock
:
7
:
LC
:
UL2601
:
NORMAL
:
OPEN
L - G
:
:
10.00mA
OFF
:
:
VOLTAGE
3.0s
:
:
OFF
0=CONT
Remote
Lock
7
:
:
LC
:
UL2601
:
NORMAL
:
OPEN
L - G
:
:
10.00mA
OFF
:
:
VOLTAGE
:
3.0s
OFF
:
Remote
Lock
EXIT
Offset
Error
Offset
Error
SETUP
Offset
Error
150799 Rev A3
Use Numeric Keys then [ENTER]
Enter the Power Low Limit or Target Frequency
[VOLTAGE]
Power Low: 0-300V, 0 = OFF
[CURRENT]
Power Low: 0-20A, 0 = OFF
[VA]
Power Low: 0-4400V, 0 = OFF
[SIMULATE]
Not Applicable
[SOURCE]
Target-F: 50 - 60Hz
Use Numeric Keys then [ENTER]
Enter theTest Time in seconds (s)
0, 0.3 - 999s
0 =Continuous
Use Function Key
Enter setup menu for scanner(s)