G6100 Plus Leakage Current Tests
The Guardian 6100 Plus Medical Safety Analyzer measures Earth Line Leakage,
Enclosure Leakage, Patient Line Leakage and Patient Auxiliary Leakage Current. The
instrument simulates the human body models listed in Table 2-8. The test modes and
other programming parameters are also listed in Table 2-8. The Earth LC Test is the sum
of all leakages of current in the DUT (device under test). The test measures the current
flowing back to Earth Ground through the ground conductor of the line (power) cord.
Device
Line Input
GB Switch
Meter
High Limit
Low Limit
Power
Power High
Power Low
Figure 2-12a through 2-15a illustrate the IEC 60601-1 circuit drawings for Earth,
Enclosure, Patient and Patient Auxiliary Leakage Current and contain multiple
abbreviations. These abbreviations used in the IEC circuit diagrams are listed herein.
Operation
www.valuetronics.com
Table 2-8: 6100 LC Test Parameters
UL544NP, UL544P, UL1563, UL2601, UL1950
Normal, Reverse, Single-Fault Normal, Single-Fault Reverse
Open, Closed
L – G, L – P2, P1 – P2
0.0001-10mA*
0 - 10mA*, 0=OFF
VOLTAGE
0 – 300V
0 – 300V
*6mA current limit for UL544NP circuit model
FE
Functional Earth
PE
Protective Earth
MD
Measuring Device
S
S
S
Single pole switch simulates the interruption of power supply conductor
1
2
3
S
S
Commutator switch reverses polarity of MAINS voltage
5
9
S
Single pole switch simulates interruption of single PE conductor
8
S
S
Switches to connect FE terminal to earthed point of measuring supply system
10
11
S
Switch to connect F-type Appied Part to earthed point of measuring supply system
12
S
Switch to connect earth to metal Accessible Part that is not an Applied Part or PE
13
T
T
1-, 2- or polyphase isolation transformers
1
2
V
V
Voltmeter including rms value
1
2
P
Socket, plug or terminal for supply connection of medical electrical equipment
1
P
Socket, plug or terminal for connection to other equipment in a medical electrical system
2
Optional Connection
1
Medical electrical equipment enclosure
Separate power supply
2
3
SIP/SOP (medical device I/O jacks other than patient leads)
4
Internal electrical power source
Patient Circuit (medical device)
5
6
Metal Accessible Part: not an Applied Part nor Protectively Earthed
Figure 2-10: IEC 60601-1 Abbreviations
CURRENT
VA
0 – 20A
0 – 4400 VA
0 – 20A
0 – 4400 VA
150799 Rev A3
SIMULATE
SOURCE
80 – 300V
80 – 300V
N/A
50-60Hz
Page 83 of 177