:TIME
:TPORT
:CHANnel (@<Box#>(<channel list>)
:OSC
:LIMIT
Interface
www.valuetronics.com
:LIMit:LOW <numeric value>
Command: :SOURce:SAFEty:STEP<n>:GB:LIMit:LOW <numeric value>
This command sets the low resistance limit for the selected step.
<numeric value> is 0 to 100 and the unit is milliohms.
:LIMit:LOW?
Command: :SOURce:SAFEty:STEP<n>:GB:LIMit:LOW?
Return Value: +1.000000E-02
This command queries the low resistance limit for the selected step.
:TEST <numeric value>
Command: :SOURce:SAFEty:STEP <n>:GB:TIME:TEST <numeric value>
This command sets the test time for the selected step.
<numeric value> is 0 to 999 and the unit is seconds.
:TIME:TEST?
Command: :SOURce:SAFEty:STEP <n>:GB:TIME:TEST?
Return Value: +6.000000E+01 (for 60 second test)
This command queries the test time for the selected step.
:TPORT 1|0
Command: :SOURce:SAFEty:STEP<n>:GB:TPORT
This command enables or disables Twin Port
Valid commands are ON or OFF 1 or 0
:TPORT?
Command: :SOURce:SAFEty:STEO<n>:GB:TPORT?
This command queries the status of Twin Port
:HIGH <channel list>
Command: :SOURce:SAFEty:STEP<n>:GB:CHAN:HIGH(@<Box#>(<channel
list>)
This command sets the high scan channels for the selected step
:HIGH?
Command: :SOURce:SAFEty:STEP<n>:GB:CHAN:HIGH?
This command queries the high scan channels for the selected step
:OPEN <numeric value>
Command: :SOURce:SAFEty:STEP<n>:OSC:LIMIT:OPEN
This command sets the Open circuit percentage
<numeric value> 0.5 to 1 (0.5 = 50%)
:SHORT <numeric value>
Command: :SOURce:SAFEty:STEP<n>:OSC:LIMIT:SHORT
This command sets the Short circuit percentage
<numeric value> 0, 1 – 5 (1 = 100%, 5 = 500%)
150799 Rev A3
Page 147 of 177
Need help?
Do you have a question about the Guardian 6100 Plus and is the answer not in the manual?