4.3.1.11
LC Power Meter Calibration
LC Voltage 300V Offset and Full
Connect an AC voltage source (Freq = 60Hz) between the L-IN terminal and the N-IN
terminal on the 6000-07 scanner.
Press [UP] to go to CAL step 5. (LCV 300V Offset
20V)
Press [STOP]
Press [START]
Using keypad enter the reading from the voltmeter followed by [ENTER].
Press [STOP] to accept reading.
Press [UP] to go to CAL step 6. (LCV 300V Full
200V)
Press [STOP]
Press [START]
Using keypad enter the reading from the voltmeter followed by [ENTER].
Press [STOP] to accept reading.
LC Current 1A Offset and Full
Connect an AC current source (Freq = 60Hz) between the L-IN terminal and the L-OUT
terminal on the 6000-07 scanner.
Press [UP] to go to CAL step 7. (LCA 1A Offset
0.08A)
Press [STOP]
Press [START]
Using keypad enter the reading from the ammeter followed by [ENTER].
Press [STOP] to accept reading.
Press [UP] to go to CAL step 8. (LCA 1A Full
0.8A)
Press [STOP]
Press [START]
Using keypad enter the reading from the ammeter followed by [ENTER].
Press [STOP] to accept reading.
Page 176 of 177
150799 Rev A3
Service & Calibration